Provide easy access for viewing the display and programming via the hand programmer.
Provide an environment suitable to the housing rating and materials of construction.
Provide a sunshield if the instrument will be mounted in direct sunlight.
Ambient temperature –40 °C to +80 °C (–40 °F to +176 °F)
Process temperature (with FKM O-ring)
Non-pressure version: –40 to +200 °C (–40 to +392 °F)
Pressure rated version: –40 to +80 °C (–40 to +176 °F)
Polypropylene lens antenna version: -40 to +80 °C (–40 to +176 °F)
See also
Auto false echo suppression (Page 55)
4.3.3 Orientation in a vessel with obstructions
Note
For best results on a vessel with obstructions, orient the front or back of the device toward the
obstructions.
Installing/mounting
4.3 Proper mounting
SITRANS LR260 with mA/HART
Operating Instructions, 11/2019, A5E32337683-AD 27