8.4.6 High Calibration Point (HIGH CALIB. PT.) (1.6.)
Distance from Sensor Reference to High Calibration Point: usually process full level.
(See Operation (Page 75) types for an illustration.)
Values Range: 0.0000 to 30.000 m
Default: 0.0 m
Related parameters Near Range (2.2.1.12.) (Page 81)
When setting the High Calibration Point value, note that echoes are ignored within Near Range
(2.2.1.12.) (Page 81)
8.4.7 Apply (1.7.)
In order to save the Quick Start settings, it is necessary to select Yes and apply changes.
Options
YES, NO, DONE
Display shows DONE when Quick Start is suc‐
cessfully completed.
Press Mode to return to Measurement mode.
8.5 Setup (2.)
Note
See Programming via the handheld programmer (Page 51) or Operating via SIMATIC PDM
(Page 59) for instructions.
Default settings in the parameter tables are indicated with an asterisk (*) unless explicitly
stated.
Values shown in the following tables can be entered via the handheld programmer
8.5.1 Device (2.1.)
8.5.2 Firmware Revision (FIRMWARE REV.) (2.1.1.)
Corresponds to the software or firmware that is embedded in the SITRANS LR260. Read only.