8.5.16.4 Up Sampling (2.2.4.2.4.)
Specifies the number of consecutive echoes that must appear above the echo currently
selected, before the measurement is accepted as valid.
Values Range: 1 to 50
Default: 5
8.5.16.5 Down Sampling (2.2.4.2.5.)
Specifies the number of consecutive echoes that must appear below the echo currently
selected, before the measurement is accepted as valid.
Values Range: 1 to 50
Default: 5
8.5.16.6 Window (2.2.4.2.6.)
A "distance window" centered on the echo
1)
, used to derive the reading. When a new
measurement is in the window, the window is re-centered and the reading is calculated.
Values Range: 0 to 30 m
Default: 0
When the value is 0, the window is automatically calculated after each measurement. 
For slower Measurement Response values, the window is narrow.
For faster Measurement Response values, the window becomes progressively wider.
Note
The echo lock window is stored as standard sample, but displayed in sensor units. Any value
entered for the echo lock window will be rounded to the nearest sample.