Operating Instructions

Table Of Contents
8.5.31 Fill Rate (FILL RATE/MIN.) (2.2.7.2.)
Defines the maximum rate at which the reported sensor value is allowed to increase. Allows you
to adjust the SITRANS LR260 response to increases in the actual material level. Fill Rate is
automatically updated whenever Response Rate is altered.
Values Range: 0 to 30 m / min.
Default: 10 m/min.
Altered by Response Rate
Related parameters Sensor Units
Enter a value slightly greater than the maximum vessel-filling rate, in Sensor Units per minute.
Options Meters/Minute
Slow 0.1
Medium 1
Fast 10
8.5.32 Empty rate (EMPTY RATE/MIN.) (2.2.7.3.)
Defines the maximum rate at which the reported sensor value is allowed to decrease. Adjusts
the SITRANS LR260 response to decreases in the actual material level. Empty Rate is
automatically updated whenever Response Rate is altered.
Values Range: 0 to 30 m / min.
Default: 10 m / min.
Altered by Response Rate
Related Sensor Units
Enter a value slightly greater than the vessel’s maximum emptying rate, in Sensor Units per
minute.
Options Meters/Minute
Slow 0.1
Medium 1
Fast 10
8.5.33 Transducer Block (TB) Values (2.2.8.)
For diagnostic purposes.
8.5.34 Level Measurement (LEVEL MEAS.) (2.2.8.2.)
The value for level. Read Only. In SIMATIC PDM, open the menu View Display, and select the
tab Display.
Parameter reference
8.5 Setup (2.)
SITRANS LR260 with mA/HART
90 Operating Instructions, 11/2019, A5E32337683-AD