Note
The last valid reading is maintained until the Fail-safe timer expires. After the timer expires, the
reading is set based on Fail-safe Material Level.
8.5.49 Fail-safe Material Level (FAILSAFE MAT. LE) (2.4.2.)
Options
HI: Use 22.6 mA (max. mA Limit) as material level
LO: Use 3.6 mA (min. mA Limit) as material level
* HOLD: Level remains at last reading
VALUE: User-selected value (defined in Fail-safe Level)
8.5.50 Fail-safe Level (2.4.4.)
Defines a user-defined level to report when the Fail-safe timer expires.
Values Range: 3.6 mA to 22.6 mA
Default: 22.6 mA
Note
This value can be used only if Fail-safe Material Level is set to VALUE.
8.6 Diagnostics (3.)
8.6.1 Echo Profile (3.1.)
Allows you to request the current echo profile either via the handheld programmer, or via
SIMATIC PDM. (For more detail see Echo Processing (Page 151).)
To request a profile via SIMATIC PDM:
1. Open the menu View – Echo Profile.
2. If desired, press Save.
3. To view the saved profile, open the menu View – Show Echo Profile.
Parameter reference
8.6 Diagnostics (3.)