Parameters
P331 Auto Record Enable
Use to enable/disable the Auto Profile Record function.
P332 Auto Record Transducer
Specifies the Transducer Point Number for which Auto Profile Records are saved.
Please note that Differential and Average operation is a MultiRanger 200 feature.
MultiRanger 200
This feature is preset to Point Number 1. (Alteration is only required if differential or
average Operation [P001 = 4 or 5] is selected.)
P333 Auto Record Interval
Enter the time to elapse after an Auto Profile Record is saved before another Auto Profile
Record can be saved (subject to all other restrictions).
Primary Index Global
Values
Range: 0 to 1
0* Off
1On
Primary Index Global
Values
Range: 0 to 2
0 Any transducer
1* Transducer One
2Transducer Two
Altered By P001 Operation = 4 or 5
Primary Index Global
Values
Range: 0.0 to 166.6 (minutes)
Preset: 120