P712 Echo Lock Sampling
The sampling criterion sets the number of consecutive echoes appearing above or below
the echo currently locked onto, that must occur before the measurements are validated
as the new reading (for Echo Lock P711 values: 1 or 2).
Example:
Setting
P711 = 2, material agitator
P712 = 5:2
Result
a new reading will not be validated unlessfive consecutive measurements higher or
two consecutive measurements lower than the current reading occur
Resetting P711 returns P712 to the respective preset values
P713 Echo Lock Window
Adjusts the size of the Echo Lock Window.
Primary Index
Single Point Model Dual Point Model
Global Transducer
Values
Range: 1:1 to 50:50
Format: x:y
x = the number of above echoes
y = the number of below echoes
Related P711 Echo Lock
P711 value P712 preset value
1, max verification 5:5
2, material agitator 5:2
Primary Index
Single Point Model Dual Point Model
Global Transducer
Values
Range: 0.000 to 99.00 m (or equivalent depending on units), nearest
acceptable value is returned
Preset: 0.000
Altered by P003 Maximum Process Speed
Related
P005 Units
P711 Echo Lock