7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 215
mmmmm
Parameters
P852 Short Shot Range
Enter the maximum distance in Units (P005) to be measured using short shot echoes.
This feature is automatically altered when Transducer (P004) is altered.
Test (P90 0 to P913)
Test Parameters are intended for use by Siemens Milltronics Service personnel.
P900 Software Revision Number
View the EPROM Rev. #.
P901 Memory
Press ENTER to activate the MultiRanger memory test.
Primary Index
Single Point Model Dual Point Model
Global Transducer
Values Range: 0 to 10 m, or equivalent depending on units
Altered By •P004 Transducer
Related
P840 Short Shot Number
P842 Short Shot Frequency
P844 Short Shot Width
P850 Short Shot Bias
P851 Short Shot Floor
Primary Index Global
Values Range: 00.00 to 99.99 (view only)
Primary Index Global
Values
Display: view only
PASS (memory test successful)
F1 RAM
F3 FLASH data
F4 FLASH code