Page 34 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06
mmmmm
Operation
Default Values
Parameter default values are indicated with an * in the parameter tables.
P000 Lock
The asterix identifies 1954 as the default value.
Parameter Reset
Returning a parameter to factory default.
1. Display the appropriate parameter number.
2. Display the appropriate index value (if required).
3. Press CLEAR .
4. Press ENTER .
Master Reset (P999)
Returns all parameters to original values.
Use Conditions:
before initial system installation
following a software upgrade
If complete reprogramming is required, use Dolphin Plus to store and retrieve parameters.
When the dual point option is enabled, P999 is indexed by transducer. Use index 00 to
reset the entire MultiRanger.
Display Readout
The following readouts are shown when the MultiRanger cannot display a number.
Primary Index Global
Value
1954 * OFF: programming permitted
-1 Simulation Controls (relays energize based on
simulated level)
other ON: Lock activated and programming not permitted
via the handheld programmer
Display Definition
Parameter has not been set
All values not same when viewing index 0
Value too large for four-digit display