Specifications
Specifications
SITRANS LR 300
Power
24–230 Vac, +15%, 40–70Hz, 28 VA (11W)
or
24–230 Vdc, +
15%, (9W)
Normal start up takes about 2 seconds to power up the unit. If something (such as
hesitation in plugging in the unit) causes a rapid power fluctuation, the unit will shut
down for 10 seconds to ensure the inrush current will not exceed specified limits.
Fuse
FU1 & FU2, Fast Acting Ceramic Bussmann ABC fuse, 1 Amp. 250V.
Interface
analog output: optically-isolated 4–20 mA into 450 max. ± 0.02 mA
accuracy
serial interface
1
:isolated
2
RS-485 (refer to RS-485 specifications)
programmer link: infrared receiver (refer to Programmer specifications on
page 10)
display (local): backlit, alphanumeric, and multi-graphic liquid crystal for
readout and entry
HART: optional Profibus PA
Notes:
Safety feature limits the inrush current during start up.
This instrument features a universal power supply. AC or DC voltage may be applied
to the same terminals.
Power consumption will vary according to voltage. Please see Appendix V, Typical
Power Consumption, on page 107.
1.
This port is used to communicate through Modbus or Dolphin Plus.
2.
Although the RS-485 is isolated, its inputs are electrically clamped to earth ground: see
page 35 for more detail.