© Siemens AG 2008 SITRANS P measuring instruments for pressure 2/2 Product overview 2/150 2/4 Transmitters for gage, absolute and differential pressure Z series for gage pressure Z series for gage and absolute pressure SITRANS P250 for differential pressure ZD series for gage and absolute pressure 2/150 2/158 2/161 2/170 2/173 2/174 2/176 2/179 2/182 2/187 2/4 2/6 2/12 2/17 2/21 2/21 2/28 2/47 2/47 2/52 2/58 Transmitters for food, pharmaceuticals and biotechnology SITRANS P Compact for gage and a
© Siemens AG 2008 SITRANS P measuring instruments for pressure Product overview ■ Overview Application 2 Description Page Software for Parameterization SITRANS P – measuring instruments for pressure, absolute pressure, differential pressure, flow and level Two- or three-wire transmitters for measuring gage and absolute pressure SITRANS P, Z series Compact single-range transmitters Analog electronics Available ex stock 2/4 – Two- or three-wire transmitters for measuring differential pressure SITRA
© Siemens AG 2008 SITRANS P measuring instruments for pressure Product overview Application Description Page Software for Parameterization 2-wire transmitter for measuing SITRANS P, MPS series (submersible sensor) 2/146 hydrostatic levels For measuring liquid levels in wells, tanks, channels, dams etc.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure Z series for gage pressure ■ Overview ■ Design The main components of the pressure transmitter are: • Brass housing with silicon measuring cell and electronics plate • Process connection • Electrical connection 2 The silicon measuring cell has a thin-film strain gage which is mounted on a ceramic diaphragm. The ceramic diaphragm can also be used for aggressive media.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure Z series for gage pressure ■ Technical specifications Design SITRANS P pressure transmitter, Z series for gage pressure Wetted parts materials Mode of operation Measuring principle ≈ 0.2 kg (≈ 0.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure Z series for gage and absolute pressure Measuring range ≥1 bar (≥14.5 psi) ■ Overview Main components: • Stainless steel housing with ceramic measuring cell and electronics module. The temperature-compensated ceramic measuring cell has a thin-film strain gage which is mounted on a ceramic diaphragm. The ceramic diaphragm can also be used for aggressive media.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure Z series for gage and absolute pressure ■ Technical specifications Power supply UH SITRANS P pressure transmitters, Z series for gage pressure, absolute pressure and level Mode of operation Terminal voltage on pressure transmitter • For current output 10 ... 36 V DC 15 ... 36 V DC • Measuring range <1 bar (<14.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure ZLWKRXW ([ SURWHFWLRQ ZLWK ([ SURWHFWLRQ ZLWKRXW ([ SURWHFWLRQ ZLWK ([ SURWHFWLRQ /HQJWK RQ YHUVLRQ IRU YROWDJH RXWSXW 9 6: 0 [ RU 137 Pressure transmitter 7MF1564-...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure Z series for gage and absolute pressure ■ Selection and Ordering data Order No. Measuring range perm. working pressure Min. Order code D) 7 MF 1 5 6 4 - 7 7 7 7 7 - 7 7 7 1 SITRANS P pressure transmitters for pressure, series Z for gage and absolute pressure 2 or 3-wire system, rising characteristic curve 777 Burst pressure 2 Max.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure Z series for gage and absolute pressure ■ Selection and Ordering data Order No. SITRANS P pressure transmitters for pressure, series Z for pressure and absolute pressure 2 or 3-wire system, rising characteristic curve Measuring range 2 Perm. working pressure min. Order code D) 7 MF 1 5 6 4 - 7 7 7 7 7 - 7 7 7 1 777 Burst pressure max.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure Z series for gage and absolute pressure ■ Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure SITRANS P250 for differential pressure ■ Overview Mode of operation p1 p2 U const. 2 U I I0, UB SITRANS P250 pressure transmitter, function diagram The piezo-resistive ceramic measuring cell (membrane) has a Wheatstone bridge circuit, on which the operating pressure P1 and P2 of the media acts at both ends.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure SITRANS P250 for differential pressure Measuring range Conditions of use Ambient conditions • Temperature of medium -15 ... +85 °C (5 ... 185 °F) Max. per- Burst missible pressure operating pressure (on either side) Max. per- Accuracy missible operating pressure (on one side) • Ambient temperature -15 ... +85 °C (5 ... 185 °F) • Storage temperature -40 ... +85 °C (-40 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure SITRANS P250 for differential pressure 1+ 1 3 2 2+ U + UB RL O 3- 2 Round connector EN 60130-9 Connection: 1 (+UB), 2 (-), 3 (+U0) UO Output voltage RL Load UB Power supply 2 + UB RL 2 1 Connection: 1 (+UB, brown), 2 (+U0, green), 3 (-, white) UO Output voltage UB Power supply RL Load Connection with voltage output 0 ... 5 V DC (0 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure SITRANS P250 for differential pressure Process connections Ø Width across flats [mm] [inch] Pipe union with screw-in nipple for outer pipe (stainless steel 1.4305/AISI 303) a X [mm] [inch] [mm] [inch] 6 a = 10 b = 12 24 65 8 a = 12 b = 14 26 67 Female thread G1/8 (stainless steel 1.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure SITRANS P250 for differential pressure ■ Selection and ordering data Order No. Order code 7 M F 1 6 4 1 - 7777 0 - 777 0 777 SITRANS P 250 pressure transmitter for differential pressure Accuracy ≤ 1 %, wetted parts ceramic/stainless steel 1.4301, scope of delivery: transmitter, mounting bracket and instruction manual, without explosion protection 2 Measuring range 0 ... 0.1 bar 0 .
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure ZD series for gage and absolute pressure ■ Overview At the rear of the housing is the electrical connection for the voltage supply using a current loop 4 ... 20 mA. The connection is made with a plug connector. At the front of the housing is the 5-digit display behind a glass cover. Underneath the display are the 3 keys for parameterizing the pressure transmitter.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure ZD series for gage and absolute pressure 2 • Upper and lower current saturation limit • Electrical damping Rated conditions The other two keys are used to set the values in the individual operating modes. • Ambient temperature -25 ... +85 °C (-13 ... +185 °F) • Storage temperature -40 ... +85 °C (-40 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure ZD series for gage and absolute pressure ■ Dimensional drawings Ø 82 (3.23) Ø 82 (3.23) M16x1,5 or ½-NPT M16x1,5 or ½-NPT SW 27 a G½B EN 837-1 ½-14 NPT G½B Type A 152 (5.98) Ø 82 (3.23) s M 2 48,5 (1.91) 48,5 (1.91) 45 (1.77) 45 (1.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure ZD series for gage and absolute pressure 2 Selection and Ordering data Order No. Ord. Code Selection and Ordering data SITRANS P pressure transmitters, ZD series for gage and absolute pressure 7M F 1 5 8 0 - Further designs Conformity error 0.25%, range adjustment 1 : 5 (max.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P Compact for gage and absolute pressure ■ Overview ■ Application The SITRANS P Compact pressure transmitter is designed for the special requirements of the food, pharmaceutical and biotechnology industries. The use of high-grade materials guarantees compliance with hygiene regulations. The SITRANS P Compact pressure transmitter is available in many versions.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P Compact for gage and absolute pressure ■ Function 2 Rated conditions The process pressure acts on a piezo-resistive semiconductor measuring bridge through a remote seal and a transmission liquid. The pressure transmitter converts the pressure values into a load-independent current. A compensation network makes the output signal largely independent of the ambient temperature.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P Compact for gage and absolute pressure Selection and Ordering data Order No. Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P Compact for gage and absolute pressure 2 Selection and Ordering data Order No. Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P Compact for gage and absolute pressure Selection and Ordering data Order No. Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P Compact for gage and absolute pressure 2 Selection and Ordering data Order No. Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P Compact for gage and absolute pressure ■ Dimensional drawings Housing Round plug with screw connection M12 Degree of protection IP65 2 For cable diam. 4 ... 10 (0.18 ... 0.39) 94 (3.7) 74,5 (2.93) 107,5 (4.23) 30 (1.18) 9 (0.35) 22,5 (0.9) » 132 (5.2) 38,5 (1.52) 74,5 (2.93) 124,5 (4.9) 74,5 (2.93) » 50 (1.97) 39 (1.53) » 8 (0.3) For cable diam. 4.5 ... 10 (0.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure ■ Overview 2 The SITRANS P300 is a digital pressure transmitter for gage and absolute pressure. All conventional thread versions are available as process connections. In addition, various hygiene-based connections and flange connections with front-flush diaphragms meet the requirements of a dead space free process connection.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure ■ Design The device comprises: • Electronics • Housing • Measuring cell ■ Function Operation of the electronics with HART communication 9 8 2 2 0 0 00 0 00 M 10 5 4 3 mC IA, UH 7 HART interface EEPROM 6 Electronics 1 Sensor Pe Perspective view of the SITRANS P300 The housing has a screw-on cover (3), with or without an inspecti
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure Operation of the electronics with PROFIBUS PA communication 9 3 4 5 mC EEPROM 6 Electronics 1 6 pe 7 Power supply unit Coupler 11 12 5 EEPROM 1 pe Input variable Power supply unit 6 6 pe 9 10 11 12 Power supply 10 Electronics BusMaster Three input keys (local operation) Digital display Power supply DP/PA coupler or lin
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure Measuring cell for gage pressure Measuring cell for gage pressure, front-flush diaphragm 1 1 6 5 5 2 4 2 4 3 pe 1 2 3 4 5 6 pe Reference pressure Measuring cell Process connection Seal diaphragm Filling liquid Relative pressure sensor Pressure as input variable Measuring cell for gage pressure, function chart The input pressure (p
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure 2 The input pressure (pe) is transferred to the absolute pressure sensor (5) via the seal diaphragm (3) and the filling liquid (4), displacing its measuring diaphragm. The displacement changes the resistance value of the four piezo resistors in the measuring diaphragm in a bridge circuit.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure Parameterization through PROFIBUS PA interface Hygiene version Fully digital communication through PROFIBUS PA, profile 3.0, is particularly user-friendly. The PROFIBUS puts the DS III PA is in connection with a process control system, e.g. SIMATIC PSC 7. Communication is possible even in a potentially explosive environment.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure ■ Technical specifications F SITRANS P300 for gage pressure and absolute pressure HART 2 PROFIBUS PA and FOUNDATION Fieldbus Gage pressure input Measured variable Gage pressure Spans (infinitely adjustable) or nominal measuring range and max. pemissible test pressure Span Max. perm. test pressure Nominal measuring range Max. perm.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure SITRANS P300 for gage pressure and absolute pressure HART PROFIBUS PA and FOUNDATION Fieldbus Input of absolute pressure, with front-flush diaphragm Measured variable Absolute pressure (front-flush) Spans (infinitely adjustable) or nominal measuring range and max. pemissible test pressure Span Max. perm.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure SITRANS P300 for gage pressure and absolute pressure HART PROFIBUS PA and FOUNDATION Fieldbus Rated operating conditions 2 Installation conditions Ambient temperature Observe the temperature class in areas subject to explosion hazard. • Measuring cell with silicone oil -40 ... +85 °C (-40 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure SITRANS P300 for gage pressure and absolute pressure HART PROFIBUS PA and FOUNDATION Fieldbus Terminal voltage on transmitter 10.5 ... 42 V DC for intrinsically safe operation: 10.5 ... 30 V DC Supplied through bus Separate power supply - Not necessary Power supply UH 2 Bus voltage • Without EEx - 9 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure HART communication HART communication 2 Communication FOUNDATION Fieldbus 230 ... 1100 Ω Protocol HART Version 5.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure Order No. Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure Order No. Selection and Ordering data 2 Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure Selection and Ordering data Order code A02 Mounting bracket made completely of stainless steel, for wall or pipe mounting Cable socket for M12 plug • Metal • Stainless steel Rating plate inscription (instead of English) • German • French • Spanish • Italian 9 9 9 9 A50 A51 FF Further designs Add "-Z" to Order No. and specify Order code.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure Selection and Ordering data Order code 2 Selection and Ordering data HART PA Further designs Add "-Z" to Order No. and specify Order code.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure ■ Dimensional drawings ෘ 2 DSSUR[ DSSUR[ + DSSUR[ + DSSUR[ DSSUR[ SITRANS P300, with oval flange, dimensions in mm (inch) ෘ DSSUR[ DSSUR[ DSSUR[ r DSSUR[ DSSUR[ DSSUR[ DSSUR[
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure ෘ + DSSUR[ 2 SITRANS P300, front-flush, dimensions in mm (inch) The diagram shows a SITRANS P300 with an example of a flange. In this drawing the height is subdivided into H1 and H2.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure Flanges to EN and ASME Sanitary process connection to DRD Flanges to EN ' + DN PN ∅D H2 25 25 40 40 50 50 80 80 40 100 40 100 16 40 16 40 115 mm (4.5“) 140 mm (5.5“) 150 mm (5.9“) 170 mm (6.7“) 165 mm (6.5“) 165 mm (6.5“) 200 mm (7.9“) 200 mm (7.9“) Approx.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for food, pharmaceuticals and biotechnology SITRANS P300 for gage and absolute pressure Thread connection G¾“, G1“ and G2“ to DIN 3852 H2 DN ¾“ 63 37 mm (1.5“) Approx. 45 mm (1.8“) 50 65 80 25 110 x ¼“ 100 25 130 x ¼“ 2“ 63 48 mm (1.9“) 63 78 mm (3.1“) Approx. 47 mm (1.9“) + ∅D + PN 1“ ' Approx. 52 mm (2“) ∅D H2 25 78 x 1/6“ 25 95 x 1/6“ Approx. 52 mm (2.1“) PN ' DN PN ∅D H2 25 40 63 mm (2.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry SITRANS P300 and DS III series with PMC connection – Technical description ■ Overview • Small long-term drift • Wetted parts made of Hastelloy • Infinitely adjustable spans from 30 mbar g to 16 bar g for DS III with HART interface • Nominal measuring range from 1 mbar g to 16 bar g for DS III with PROFIBUS PA and FOUNDATION Fieldbus interface • Infinitely adjustable spans from 30 mbar g to
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry SITRANS P300 and DS III series with PMC connection – Technical description ■ Design SITRANS P300 SITRANS P, DS III series 2 The device comprises: • Electronics • Housing • Measuring cell Device front view, SITRANS P DS III The transmitter comprises a range of different components, depending on the order specifications. The various options are listed in the ordering information.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry SITRANS P300 and DS III series with PMC connection – Technical description ■ Function Operation of the electronics with PROFIBUS PA communication Operation of the electronics with PROFIBUS PA communication 9 8 M 2 3 4 mC 5 mC 10 5 4 3 2 2 7 PA interface IA, UH EEPROM 7 6 HART interface Electronics 6 1 Electronics 1 Pe 1 2 3 4 5 6 7 8 9 10 IA UH Pe Sensor EEPROM 6
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry SITRANS P300 and DS III series with PMC connection – Technical description Mode of operation of the FOUNDATION Fieldbus electronics 0 0 00 0 00 M 2 2 Measuring cell for gage pressure with front-flush diaphragm 9 8 Mode of operation of the measuring cell 3 4 5 mC EEPROM 7 FF interface Power supply 10 5 4 1 Power supply unit 6 Coupler Electronics 1 pe 1 2 3 4 5 6 7 EEPRO
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry SITRANS P300 and DS III series with PMC connection – Technical description Parameterization through PROFIBUS PA interface 230 ... 500 W SITRANS P transmitter Power supply PC or laptop HART modem Fully digital communication through PROFIBUS PA, profile 3.0, is particularly user-friendly. The PROFIBUS puts the DS III PA is in connection with a process control system, e.g. SIMATIC PSC 7.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry DS III series with PMC connection ■ Technical specifications SITRANS P, DS III series for gage pressure with PMC connection for the paper industry HART 2 PROFIBUS PA or FOUNDATION Fieldbus Input Measured variable Gage pressure Spans (infinitely adjustable) or nominal measuring range and max. permissible test pressure Span Max. perm. test pressure Nominal measuring range Max. perm.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry DS III series with PMC connection SITRANS P, DS III series for gage pressure with PMC connection for the paper industry HART PROFIBUS PA or FOUNDATION Fieldbus Rated operating conditions Degree of protection (to EN 60529) IP65, IP68, NEMA X, enclosure cleaning, resistant to lyes, steam to 150° C (302 °F) Process temperature -20 ... +100 °C (-4 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry DS III series with PMC connection SITRANS P, DS III series for gage pressure with PMC connection for the paper industry Communication FOUNDATION Fieldbus Function blocks HART communication 2 HART communication 230 ... 1100 Ω Protocol HART Version 5.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry DS III series with PMC connection Selection and Ordering data Order No. Selection and Ordering data SITRANS P pressure transmitters for gage pressure, with PMC connection series DS III HART 7MF 4 1 3 3 - SITRANS P pressure transmitter for gage pressure, with PMC connection Measuring cell filling Silicone oil Inert liquid Measuring cell cleaning Standard Grease-free Span 0.01 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry DS III series with PMC connection Selection and Ordering data Order code Selection and Ordering data HART PA Further designs FF Add "-Z" to Order No. and specify Order code. 2 Order code HART PA Additional data FF Add "-Z" to Order No. and specify Order code. A50 9 9 9 Measuring range to be set Specify in plain text (max. 5 digits): Y01: ... up to ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry DS III series with PMC connection ■ Dimensional drawings 144 (567) Ø 80 (3.15) 20 (0.8) 74 (2.9) 1 2 3 4 5 6 7 1) 2) 7 H2 1 2 2 Space for rotation of housing 90 (3.6)2) 6 3 4 1) 5 1) 100 (3.94) 50 (1.97) approx. 30 (1.18) 51 (2.0) H1 = approx. 130 (5.12) 15 (0.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry SITRANS P300 with PMC connection ■ Technical specifications SITRANS P300 for gage pressure with PMC connection for the paper industry HART 2 PROFIBUS PA Input Measured variable Gage pressure (flush-mounted) Spans (infinitely adjustable) or nominal measuring range and max. permissible test pressure Span Max. perm. test pressure Nominal measuring range Max. perm. test pressure 0.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry SITRANS P300 with PMC connection SITRANS P300 for gage pressure with PMC connection for the paper industry HART PROFIBUS PA Medium conditions Process temperature • Measuring cell with silicone oil 2 -40 ... +100 °C (-40 ... +212 °F) Design Weight (without options) Approx. 1 kg (2.2 lb) Housing material Stainless steel, mat. No. 1.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry SITRANS P300 with PMC connection SITRANS P300 for gage pressure with PMC connection for the paper industry Communication FOUNDATION Fieldbus Function blocks HART communication 2 HART communication 230 ... 1100 Ω Protocol HART Version 5.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry SITRANS P300 with PMC connection Order No. Selection and Ordering data Further designs 4 ... 20 mA/HART F) 7 M F 8 1 2 3 - PROFIBUS PA F) 7 M F 8 1 2 4 - FOUNDATION Fieldbus (FF) F) 7 M F 8 1 2 5 - Measuring cell filling Silicone oil Inert liquid Span 1 bar g 4 bar g 16 bar g Measuring cell cleaning Standard Cleanliness level 2 to DIN 25410 (14.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage pressure for the paper industry SITRANS P300 with PMC connection ■ Dimensional drawings ෘ + $SSUR[ + $SSUR[ 2 r SITRANS P300 pressure transmitters for gage pressure, with PMC connection, dimensions in mm (inch) The diagram shows a SITRANS P300 with an example of a flange. In this drawing the height is subdivided into H1 and H2.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III, DS III PA and DS III FF series Technical description ■ Overview 2 SITRANS P pressure transmitters, DS III series, are digital pressure transmitters featuring extensive user-friendliness and high accuracy. The parameterization is performed using control keys, over HART communication, PROFIBUS-PA or FOUNDATION Fieldbus interface.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III, DS III PA and DS III FF series Technical description 2 Pressure transmitter for gage pressure • Measured variable: Gage pressure of aggressive and non-aggressive gases, vapors and liquids. • Span (infinitely adjustable) for DS III HART: 0.01 ... 400 bar g (0.145 ... 5802 psi g) • Nominal measuring range for DS III PA and FF: 1 ... 400 bar g (14.5 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III, DS III PA and DS III FF series Technical description ■ Function Operation of the electronics with PROFIBUS PA communication Operation of the electronics with PROFIBUS PA communication 9 8 M 2 3 4 mC 5 mC 10 5 4 3 2 2 7 PA interface IA, UH EEPROM 7 6 HART interface Electronics 6 1 Electronics 1 Pe 1 2 3 4 5 6 7 8 9 10 IA UH Pe Sensor EEP
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III, DS III PA and DS III FF series Technical description Mode of operation of the FOUNDATION Fieldbus electronics 0 0 00 0 00 M 2 2 Measuring cell for gage pressure 9 8 Mode of operation of the measuring cells 3 4 5 mC EEPROM 7 FF interface Power supply 10 5 4 3 1 Power supply unit 6 Coupler Electronics 1 pe 1 2 3 4 5 6 7 EEPROM p e Measurin
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III, DS III PA and DS III FF series Technical description Measuring cell for absolute pressure from gage pressure series Measuring cell for differential pressure and flow 2 3 4 1 2 3 4 5 6 7 5 4 1 3 1 1 2 3 4 5 pe 1 5 + Measuring cell Process connection Seal diaphragm Filling liquid Silicon absolute pressure sensor Absolute pressure as input variable 7 6
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III, DS III PA and DS III FF series Technical description of the measuring cell (7), thus protecting the silicon pressure sensor from overloads. Parameterization DS III 2 Depending on the version, there are a range of options for parameterizing the pressure transmitter and for setting or scanning the parameters.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III, DS III PA and DS III FF series Technical description Diagnostic functions for DS III PA and FF • Event counter • Slave pointer • Maintenance timer • Simulation functions • Display of zero correction • Limit transmitter • Saturation alarm 2 Physical dimensions available for the display Physical variable Physical dimensions Pressure (setting can also be made in
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage pressure ■ Technical specifications SITRANS P, DS III series for gage pressure HART 2 PROFIBUS PA or FOUNDATION Fieldbus Input Measured variable Gage pressure Spans (infinitely adjustable) or nominal measuring range and max. permissible test pressure Span Max. perm. test pressure Nominal measuring range Max. perm. test pressure 0.01 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage pressure SITRANS P, DS III series for gage pressure HART PROFIBUS PA or FOUNDATION Fieldbus Rated operating conditions Degree of protection (to EN 60529) IP65 2 Process temperature • Measuring cell with silicone oil filling -40 ... +100 °C (-40 ... +212 °F) • Measuring cell with inert filling liquid -20 ... +100 °C (-4 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage pressure SITRANS P, DS III series for gage pressure HART PROFIBUS PA or FOUNDATION Fieldbus Certificate and approvals 2 Classification according to pressure equipment directive (DRGL 97/23/EC) For gases of fluid group 1 and liquids of fluid group 1; complies with requirements of Article 3, paragraph 3 (sound engineering practice) Explosion pro
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage pressure HART communication HART communication Communication FOUNDATION Fieldbus 230 ... 1100 Ω Protocol HART Version 5.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage pressure Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage pressure Selection and Ordering data Order No. Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage pressure Selection and Ordering data Order code 2 Selection and Ordering data HART PA Further designs Add "-Z" to Order No. and specify Order code.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage pressure ■ Dimensional drawings 55 (2.2) approx 30 (1.2) 4) 100 (3.94) 28 (1.1) 6 20 (0.8) 117 (4.6) ø 80 (3.15) 74 (2.9) 51 (2.0) 3 1) 8 4 120 (4.7) 68 (2.7) 72 (2.8) 1 36,5 (1,44) 1) 5 2 2 171 (6.7) 237 (9.3) 144 (5.67) 50 (1.97) 15 (0.6) 123 (4.8) 7 5) min. 90 (3.6) 1 2 3 4 5 6 7 8 105 (4.1) ø 50 ... 60 (1.97 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage pressure 4) ca. 30 (1.2) 20 (0.8) 117 (4.6) ø 80 (3.15) 100 (3.94) ca. 96 (3.78) 28 (1.1) 6 3 3 1) 8 4 120 (4.7) 68 (2.7) 72 (2.8) 1 36,5 (1,44) 1) 5 2 171 (6.7) 237 (9.3) 51 (2.0) 74 (2.9) 2 55 (2.2) 144 (5.67) 50 (1.97) 15 (0.6) 123 (4.8) 7 5) min. 90 (3.6) 1 2 3 4 5 6 7 8 105 (4.1) ø 50 ... 60 (1.97 ... 2.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage and absolute pressure, with front-flush diaphragm ■ Technical specifications SITRANS P, DS III series for gage and absolute pressure, with front-flush diaphragm HART PROFIBUS PA or FOUNDATION Fieldbus Input gage pressure, with front-flush diaphragm 2 Measured variable Gage pressure, flush-mounted Spans (infinitely adjustable) or nominal measu
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage and absolute pressure, with front-flush diaphragm SITRANS P, DS III series for gage and absolute pressure, with front-flush diaphragm HART PROFIBUS PA or FOUNDATION Fieldbus Influence of ambient temperature 2 • at -10 ... +60 °C (14 ... 140 °F) ≤ (0.1 ⋅ r + 0.2)% • at -40 ... -10 °C and +60 ... +85 °C (-40 ... +14 °F and 140 ... 185 °F) ≤ (0.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage and absolute pressure, with front-flush diaphragm SITRANS P, DS III series for gage and absolute pressure, with front-flush diaphragm HART PROFIBUS PA or FOUNDATION Fieldbus Supplied through bus Power supply UH Terminal voltage on transmitter 10.5 ... 45 V DC 10.5 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage and absolute pressure, with front-flush diaphragm HART communication HART communication 2 Communication FOUNDATION Fieldbus 230 ... 1100 Ω Protocol HART Version 5.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage and absolute pressure, with front-flush diaphragm Selection and Ordering data Order No. Measuring cell filling Silicone oil Inert liquid FDA compliant fill fluid • Neobee oil Measuring cell cleaning Standard Grease-free 1 3 Standard 4 B C D E 13 ... 1300 mbar a2) 0.05 ... 5 bar a2) 3 ... 30 bar a2) (0.19 ... 18.9 psi a)2) (0.7 ... 72.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage and absolute pressure, with front-flush diaphragm Selection and Ordering data Order code Selection and Ordering data HART PA Further designs FF Add "-Z" to Order No. and specify Order code.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage and absolute pressure, with front-flush diaphragm Selection and Ordering data Order code Order code Selection and Ordering data HART PA Further designs FF HART PA Additional data FF Add "-Z" to Order No. and specify Order code. Add "-Z" to Order No. and specify Order code.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage and absolute pressure, with front-flush diaphragm ■ Dimensional drawings DSSUR[ 6SDFH IRU URWDWLRQ RI KRXVLQJ 90 (3.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage and absolute pressure, with front-flush diaphragm Thread connection G¾“, G1“ and G2“ to DIN 3852 DN PN ∅D H2 DN PN ∅D H2 50 40 105 mm (4.1“) Approx. 52 mm (2“) ¾“ 63 37 mm (1.5“) Approx. 45 mm (1.8“) 1“ 63 48 mm (1.9“) Approx. 47 mm (1.9“) 2“ 63 78 mm (3.1“) Approx.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for gage and absolute pressure, with front-flush diaphragm DN PN ∅D H2 50 25 78 x 1/6“ 65 25 95 x 1/6“ Approx. 52 mm (2.1“) 80 25 110 x ¼“ 100 25 130 x ¼“ ' Aseptic flange with notch to DIN 11864-2 Form A + DN ' PN ∅D H2 Approx. 52 mm (2.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from gage pressure series) ■ Technical specifications SITRANS P, DS III series for absolute pressure (from the gage pressure series) HART PROFIBUS PA or FOUNDATION Fieldbus Input 2 Measured variable Absolute pressure Spans (infinitely adjustable) or nominal measuring range and max. permissible test pressure Span Max. perm.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from gage pressure series) SITRANS P, DS III series for absolute pressure (from the gage pressure series) HART PROFIBUS PA or FOUNDATION Fieldbus Rated operating conditions 2 Degree of protection (to EN 60529) IP65 Process temperature • Measuring cell with silicone oil filling -40 ... +100 °C (-40 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from gage pressure series) SITRANS P, DS III series for absolute pressure (from the gage pressure series) HART PROFIBUS PA or FOUNDATION Fieldbus Certificate and approvals Classification according to pressure equipment directive (DRGL 97/23/EC) For gases of fluid group 1 and liquids of fluid group 1; complies with requirements of Ar
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from gage pressure series) HART communication HART communication 2 Communication FOUNDATION Fieldbus 230 ... 1100 Ω Protocol HART Version 5.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from gage pressure series) Selection and Ordering data Order No. Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from gage pressure series) Selection and Ordering data Order No. 2 Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from gage pressure series) Selection and Ordering data Order code Order code Selection and Ordering data HART PA Further designs FF Additional data HART PA FF Add "-Z" to Order No. and specify Order code. Add "-Z" to Order No. and specify Order code.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from gage pressure series) ■ Dimensional drawings 117 (4.6) ø 80 (3.15) 20 (0.8) 74 (2.9) 4) ca. 30 (1.2) 100 (3.94) ca. 96 (3.78) 28 (1.1) 6 3 3 1) 8 4 1) 72 (2.8) 1 120 (4.7) 68 (2.7) 5 2 171 (6.7) 237 (9.3) 51 (2.0) 36,5 (1,44) 2 55 (2.2) 144 (5.67) 50 (1.97) 15 (0.6) 123 (4.8) 7 5) min. 90 (3.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from gage pressure series) 55 (2.2) approx 30 (1.2) 4) 100 (3.94) 28 (1.1) 6 20 (0.8) 3 1) 8 4 120 (4.7) 68 (2.7) 72 (2.8) 1 36,5 (1,44) 1) 5 2 2 117 (4.6) ø 80 (3.15) 74 (2.9) 51 (2.0) 171 (6.7) 237 (9.3) 144 (5.67) 50 (1.97) 15 (0.6) 123 (4.8) 7 5) min. 90 (3.6) 1 2 3 4 5 6 7 8 105 (4.1) ø 50 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from differential pressure series) ■ Technical specifications SITRANS P, DS III series for absolute pressure (from differential pressure series) HART 2 PROFIBUS PA or FOUNDATION Fieldbus Input Measured variable Absolute pressure pressure Spans (infinitely adjustable) or nominal measuring range and max.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from differential pressure series) SITRANS P, DS III series for absolute pressure (from differential pressure series) HART PROFIBUS PA or FOUNDATION Fieldbus Rated operating conditions Degree of protection (to EN 60529) IP65 2 Process temperature • Measuring cell with silicone oil filling -40 ... +100 °C (-40 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from differential pressure series) SITRANS P, DS III series for absolute pressure (from differential pressure series) HART PROFIBUS PA or FOUNDATION Fieldbus Certificate and approvals 2 Classification according to pressure equipment directive (DRGL 97/23/EC) For gases of fluid group 1 and liquids of fluid group 1; complies with re
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from differential pressure series) HART communication HART communication Communication FOUNDATION Fieldbus 230 ... 1100 Ω Protocol HART Version 5.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from differential pressure series) Order No. Selection and Ordering data 2 Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from differential pressure series) Order No. Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from differential pressure series) Selection and Ordering data Order code 2 Selection and Ordering data HART PA Further designs FF Order code HART PA Further designs FF Add "-Z" to Order No. and specify Order code. Add "-Z" to Order No. and specify Order code.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from differential pressure series) ■ Dimensional drawings 3 approx. 261 (10.3) 9 134 (5.27) 1 + Space for rotation of housing 10 8 7 11 ø 50 ... 60 (1.97 ... 2.36) approx. 96 (3.78) 3 72 (2.83) 105 (4.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for absolute pressure (from differential pressure series) approx. 30 (1.2) 2 6 approx. 261 (10.3) 9 1 + Space for rotation of housing 10 8 7 52 (2.05) ø 50 ... 60 (1.97 ... 2.36) 11 96 (3.8) 1) 4 3 198 (7.8) 117 (4.6) Ø 80 (3.15) 5 1) 100 (3.94) 50 (1.97) 51 (2.0) 69 (2.7) 90 (3.6)2) 134 (5.27) 2 144 (5.67) 20 (0.8) 15 (0.6) 72 (2.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow ■ Technical specifications SITRANS P, DS III series, for differential pressure and flow HART PROFIBUS PA or FOUNDATION Fieldbus Input Measured variable Differential pressure and flow Spans (infinitely adjustable) or nominal measuring range and max.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow SITRANS P, DS III series, for differential pressure and flow HART 2 PROFIBUS PA or FOUNDATION Fieldbus ≤ 0,2 • Square-root characteristic (flow 25 ... 50%) - r ≤ 10 ≤ 0,2% - - 10 < r ≤ 30 ≤ 0,4% - Long-term drift (temperature change ± 30 °C ≤ (0.25 ⋅ r)% every 5 years (± 54 °F)) static pressure max.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow SITRANS P, DS III series, for differential pressure and flow HART PROFIBUS PA or FOUNDATION Fieldbus Terminal voltage on transmitter 10.5 ... 45 V DC 10.5 ... 30 V DC in intrinsically-safe mode - Separate 24 V power supply necessary - No Supplied through bus Power supply UH 2 Bus voltage • Not Ex - 9 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow SITRANS P, DS III series, for differential pressure and flow HART PROFIBUS PA or FOUNDATION Fieldbus Certificate and approvals 2 Classification according to pressure equipment directive (DRGL 97/23/EC) PN 32/160 (MWP 464/2320 psi) For gases of fluid group 1 and liquids of fluid group 1; complies with requirements of Ar
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow HART communication HART communication Communication FOUNDATION Fieldbus 230 ... 1100 Ω Protocol HART Version 5.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow 2 Selection and Ordering data Order No. Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow Selection and Ordering data Order No. Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow Selection and Ordering data Order code 2 Selection and Ordering data HART PA Further designs Add "-Z" to Order No. and specify Order code.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow Selection and Ordering data Order code HART PA Additional data Add "-Z" to Order No. and specify Order code. Measuring range to be set Specify in plain text: • With linear characteristic (max. 5 digits): Y01: ... up to ... mbar, bar, kPa, MPa, psi • With square-rooted characteristic (max. 5 digits): Y02: ... up to ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow 2 Selection and Ordering data Order No. Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow Selection and Ordering data Order No. Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow Selection and Ordering data Order code 2 Selection and Ordering data HART PA Further designs FF Order code HART PA Further designs FF Add "-Z" to Order No. and specify Order code. Add "-Z" to Order No. and specify Order code.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow ■ Dimensional drawings 3 approx. 261 (10.3) 9 134 (5.27) 1 + Space for rotation of housing 10 8 7 11 ø 50 ... 60 (1.97 ... 2.36) approx. 96 (3.78) 52 (2.05) 3 72 (2.83) 105 (4.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow approx. 30 (1.2) 2 6 approx. 261 (10.3) 9 1 + Space for rotation of housing 10 8 7 52 (2.05) ø 50 ... 60 (1.97 ... 2.36) 11 96 (3.8) 1) 4 3 198 (7.8) 117 (4.6) Ø 80 (3.15) 5 1) 100 (3.94) 50 (1.97) 51 (2.0) 69 (2.7) 90 (3.6)2) 134 (5.27) 2 144 (5.67) 20 (0.8) 15 (0.6) 72 (2.83) 105 (4.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow 144 (5.67) 3 6 2 3 9 approx. 85 (3.35)6) Space for rotation of housing 90 (3.6)4) 100 (3.94) 2 4 1) 5 1) approx. 30 (1.18)8) 8 + approx. 217 (8.54)7) Ø 80 (3.15) 74 (2.9) approx. 96 (3.78) 50 (1.97) 51 (2.0) 20 (0.8) 15 (0.6) 8 1 67 (2.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow 144 (5.67) 90 (3.6)4) 6 50 (1.97) 2 9 approx. 85 (3.35)6) 5 1) 100 (3.94) 3 4 1) Space for rotation of housing 2 diam. 80 (3.15) 74 (2.9) approx 30 (1.18)6) 20 (0.8) 51 (2.0) 8 + approx. 217 (8.54)7) 15 (0.6) 8 1 67 (2.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for differential pressure and flow 2 90 (3.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for level ■ Technical specifications SITRANS P, DS III series for level HART 2 PROFIBUS PA or FOUNDATION Fieldbus Input Measured variable Level Spans (infinitely adjustable) or nominal measuring range and max. permissible working pressure Span Maximum working pressure Nominal measuring range Maximum working pressure 25 ... 250 mbar g (0.36 ... 3.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for level SITRANS P, DS III series for level HART PROFIBUS PA or FOUNDATION Fieldbus - 250 mbar (3.63 psi) measuring cell ≤ (0.3 ⋅ r)% per nominal pressure ≤ 0.3% per nominal pressure - 600 mbar (8.7 psi) measuring cell ≤ (0.15 ⋅ r)% per nominal pressure ≤ 0.15% per nominal pressure - 1600 and 5000 mbar (23.2 and 72.5 psi) measuring cells ≤ (0.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for level SITRANS P, DS III series for level HART PROFIBUS PA or FOUNDATION Fieldbus Certificate and approvals 2 Classification according to pressure equipment For gases of fluid group 1 and liquids of fluid group 1; complies with requirements of Article 3, directive (DRGL 97/23/EC) paragraph 3 (sound engineering practice) Explosion protection • Intrinsi
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for level HART communication HART communication Communication FOUNDATION Fieldbus 230 ... 1100 Ω Protocol HART Version 5.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for level 2 Selection and Ordering data Order No. 1) SITRANS P pressure transmitters for level series DS III HART 7MF 4 6 3 3 - 2) 77 Y77 - 7777 3) Measuring cell filling Silicone oil Measuring cell cleaning Standard Span 25 ... 250 mbar 25 ... 600 mbar 53 ... 1600 mbar 0.16 ... 5 bar (0.363 ... 3.63 psi) (0.363 ... 8.70 psi) (0.77 ... 23.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for level Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for level Selection and Ordering data Order code Additional data HART PA FF Add "-Z" to Order No. and specify Order code. 2 9 Measuring range to be set Specify in plain text (max. 5 digits): Y01: ... up to ... mbar, bar, kPa, MPa, psi Y01 Measuring point number (TAG No.) Max. 16 characters, specify in plain text: Y15: ..............................
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for level Selection and Ordering data Order code HART PA and FF Further designs Add "-Z" to Order No. and specify Order code. A01 Flame flashover lock-out For mounting on zone 0 (including documentation) 9 9 Quality inspection certificate (Factory cali- C11 bration) to IEC 60770-2 9 9 Acceptance test certificate To EN 10204-3.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for level ■ Dimensional drawings approx. 30 (1.18)5) 185 100 (3.94) 2 51 (2.0) 6 15 (0.6) 144 (5.67) 74 (2.9) 51) 90 (3.6)4) L 9 165 (6.5) 2 3 3 1 k D d5 d4 8 108 (4.25) 54 (2.1) Space for rotation of housing nxd 2 b approx. 96 (3.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for gage, absolute and differential pressure, flow and level DS III series for level approx. 30 (1.18) 185 100 (3.94) 51 (2.0) 6 15 (0.6) 144 (5.67) 74 (2.9) 2 3 51) 90 (3.6)4) L 9 165 (6.5) 2 1 k D d5 d4 8 108 (4.25) 54 (2.
© Siemens AG 2008 SITRANS P measuring instruments for pressure SITRANS P Accessories Supplementary electronics for 4-wire connection ■ Overview Structural design 2 Dimensions (W x H x D) in mm (inch) 80 x 120 x 60 (3.15 x 4.72 x 2.36) Electrical connection Screw terminals (Pg 13.5 cable inlet) or Han 7D / Han 8U plug Power supply Supply voltage 230 V AC (-10 ... +6%, 47 ... 63 Hz, approx. 6 VA) or 24 V AC/DC (24 V AC ± 10%, 47 ... 63 Hz, approx.
© Siemens AG 2008 SITRANS P measuring instruments for pressure SITRANS P Accessories Supplementary electronics for 4-wire connection Selection and Ordering data Order code Supplementary electronics for 4-wire connection Order No. of the transmitter 7MF4.33-.....-1AB. add "-Z" and Order code. V77 Power supply 24 V AC/DC 230 V AC Electrical connection Terminals; 2 Pg screwed glands, to left 2 Han 7D/Han 8U plugs incl. mating connector, to left 1 Han 7D plug incl.
© Siemens AG 2008 SITRANS P measuring instruments for pressure SITRANS P Accessories Accessories / spare parts for SITRANS P P300 and DS III series 2 Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure SITRANS P Accessories Accessories / spare parts for SITRANS P P300 and DS III series Order No. Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure SITRANS P Accessories Accessories / spare parts for SITRANS P P300 and DS III series Selection and Ordering data Order No. Spare parts / Accessories 2 Replacement measuring cell for differential pressure and PN 420 (MWP 6092 psi) for SITRANS P, DS III, DS III PA and DS III FF series Measuring cell filling Silicone oil Measuring cell cleaning Standard Measured span 2.5 ... 250 mbar 6 ... 600 mbar 16 ... 1600 mbar 50 ... 5000 mbar 0.3 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure SITRANS P Accessories Accessories / spare parts for SITRANS P P300 and DS III series Order No. Selection and Ordering data Spare parts / Accessories Mounting bracket and mounting parts for pressure transmitters DS III, DS III PA and DS III FF series (7MF403.-.....-..C.) for absolute pressure transmitters DS III, DS III PA and DS III FF series (7MF423.-.....-..C.
© Siemens AG 2008 SITRANS P measuring instruments for pressure SITRANS P Accessories Accessories / spare parts for SITRANS P P300 and DS III series Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure SITRANS P Accessories Accessories / spare parts for SITRANS P P300 and DS III series ■ Dimensional drawings 2 5 5 Mounting bracket for SITRANS P gage and absolute pressure transmitter, DS III series Material of mounting bracket: Sheet-steel Mat. No. 1.0330, chrome-plated, or stainless steel Mat. No. 1.
© Siemens AG 2008 SITRANS P measuring instruments for pressure SITRANS P Accessories Factory-mounting of valve manifolds on SITRANS P transmitters ■ Overview 2 The 7MF9011-4FA valve manifolds are sealed with PTFE sealing tape between the transmitter and the valve manifold.
© Siemens AG 2008 SITRANS P measuring instruments for pressure SITRANS P Accessories Factory-mounting of valve manifolds on SITRANS P transmitters ■ Dimensional drawings Valve manifolds mounted on SITRANS P DS III 2 Valve manifold 7MF9011-4EA with mounted relative pressure and absolute pressure transmitters, dimensions in mm (inch) Valve manifold 7MF9011-4FA with mounted relative pressure and absolute pressure transmitters Val
© Siemens AG 2008 SITRANS P measuring instruments for pressure SITRANS P Accessories Factory-mounting of valve manifolds on SITRANS P transmitters 2 Valve manifold 7MF9411-5CA with mounted differential pressure transmitter Valve manifold 7MF9411-5BA with mounted differential pressure transmitter 5 Valve manifold 7MF9411-5BA with mounted differential pressure transmitter, dimensions in mm (inch) 5 Valve manifold 7MF9411-5CA with mounted d
© Siemens AG 2008 SITRANS P measuring instruments for pressure SITRANS P Accessories Factory-mounting of valve manifolds on SITRANS P transmitters Valve manifolds mounted on SITRANS P300 Valve manifold 7MF9011-4EA with mounted relative pressure and absolute pressure transmitters, dimensions in mm (inch) Valve manifold 7MF9011-4FA with mounted relative pressure and absolute pressure transmitters Valve manifold 7MF9011-4EA with mounted relative pressure
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for hydrostatic level MPS series (submersible sensor) ■ Overview Sensor Diaphragm 2 U const. p U I EM +2 (brown) I , U -1 (blue) 0 B Vent pipe Screen SITRANS P pressure transmitters, MPS series, mode of operation and wiring diagram SITRANS P pressure transmitters, MPS series, are submersible sensors for hydrostatic level measurements.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for hydrostatic level MPS series (submersible sensor) Long-term stability Zero and span • 1 ... 6 mH2O (3 ... 18 ftH2O) 0.25% of full-scale value/year • ≥ 6 mH2O (≥ 18 ftH2O) 0.2% of full-scale value/year Rated operating conditions Ambient conditions • Process temperature -10 ... +80 °C (+14 ... +176 °F) • Storage temperature -40 ... +100 °C (-40 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for hydrostatic level MPS series (submersible sensor) Order No. C) 7 M F 1 5 7 0 - 7 7 A 0 7 7 7 7 SITRANS P pressure transmitters for pressure, MPS series (submersible sensor) Selection and Ordering data Protective cap Cable sheath + (brown) 2-wire system Cable length L 10 m 10 m 25 m 25 m 25 m 25 m 0 ... 6 ftH2O 0 ... 12 ftH2O 0 ... 18 ftH2O 0 ... 30 ftH2O 0 ... 60 ftH2O 32 ft 32 ft 82 ft 82 ft 82 ft } } } } } } 158 (6.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Transmitters for hydrostatic level MPS series (submersible sensor) ■ More information Determination of the measuring range in case of media with a density ≠ 1000 kg/m3 (medium ≠ water) 2 + ρ Calculation of the measuring range: p=ρxgxH with: ρ = density of medium g = local acceleration due to gravity H = maximum level Example: Medium: Diesel fuel = 850 kg/m3 Acceleration due to gravity: 9.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Remote seals for transmitters and pressure gages Technical description ■ Overview capillaries of equal length must always be used to connect a remote seal to a pressure transmitter for differential pressure. In many cases the pressure transmitter and the measured medium have to be physically separated. It is then necessary to use a remote seal.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Remote seals for transmitters and pressure gages Technical description ■ Function The measured pressure is transferred from the diaphragm to the filling liquid and passes through the capillary to the measuring chamber of the pressure transmitter. The interior of the diaphragm seal and of the capillary, as well as the measuring chamber of the transmitter, are filled gas-free by the filling liquid.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Remote seals for transmitters and pressure gages Technical description Response time 2 The response time is dependent on the following factors: • Internal diameter of the capillary: The bigger the internal diameter, the shorter the response time • Viscosity of the filling liquid The greater the viscosity, the longer the response time • Length of the capillary: The longer the capillary, the longer the response time • Pressure in the pressure m
© Siemens AG 2008 SITRANS P measuring instruments for pressure Remote seals for transmitters and pressure gages Technical description ■ Technical specifications Temperature error Diaphragm seals Temperature errors of diaphragm seals when connected to pressure transmitters for pressure, absolute pressure, differential pressure (single-sided) and level Nominal diameter/ design Diaphragm diameter mm Sandwich design or with flange to EN 1092-1 Sandwich design or with flange to ASME B16.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Remote seals for transmitters and pressure gages Technical description Temperature errors of diaphragm seals with connection to differential pressure transmitters (double-sided) Nominal diameter/ design Diaphragm diameter Temperature error Temperature error of cap- Temperature error of remote seal illary of process flange/connection spigot mm (inch) mbar/ 10 K (psi/ 10 K) DN 50 without tube 59 (2.32) 0.3 (0.0043) 0.3 (0.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Remote seals for transmitters and pressure gages Technical description Temperature error Clamp-on seals Temperature errors of clamp-on seals when connected to pressure transmitters for pressure and absolute pressure, and with singlesided connection to pressure transmitters for differential pressure Nominal diameter/ design Temperature error of remote seal Temperature error of capillary Temperature error of process flange/connection spigot R
© Siemens AG 2008 SITRANS P measuring instruments for pressure Remote seals for transmitters and pressure gages Technical description Calculation of the temperature error Dependence of temperature error on diaphragm material The following equation is used to calculate the temperature error: The temperature errors listed in the previous table are based on the use of stainless steel as the diaphragm material.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Remote seals for transmitters and pressure gages Technical description Response times The values listed in the following table are the response times (in seconds per meter of capillary) for a change in pressure which corresponds to the set span. The response times are independent of the set span within the range of the respective transmitter. The response times are of insignificant importance for spans above 10 bar (145 psi).
© Siemens AG 2008 SITRANS P measuring instruments for pressure Diaphragm seals of sandwich design For gage, absolute, differential pressure and flow with flexible capillary ■ Overview 2 Diaphragm seals of sandwich design ■ Technical specifications Diaphragm seals of sandwich design Maximum pressure See above and the technical data of the pressure transmitters Tube length Without tube as standard (tube available on request) Nominal diameter Nominal pressure • DN 50 PN 16 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Diaphragm seals of sandwich design For gage, absolute, differential pressure and flow with flexible capillary Selection and Ordering data Order No. Ord.code Order No. Ord.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Diaphragm seals of sandwich design For gage, absolute, differential pressure and flow with flexible capillary ■ Dimensional drawings 2 D dM D dM b b _ R ³ (5. 150 9) R ³ (5. 150 9) l l + Connection to EN 1092-1 Nom. diam. Nom. press. b D dM l mm mm mm mm Connection to EN 1092-1 Nom. diam. Nom. press. b D dM l DN 50 PN 16 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Diaphragm seals of flange design For gage, absolute, differential pressure and flow with flexible capillary ■ Overview 2 Diaphragm seals of flange design ■ Technical specifications Diaphragm seals of flange design with flexible capillary Sealing material in the process flanges Nominal diameter Nominal pressure • DN 50 (recommendable only for pressure transmitters for pressure) PN 10 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Diaphragm seals of flange design For gage, absolute, differential pressure and flow with flexible capillary Selection and Ordering data 2 Order No. Ord. code Order No. Ord.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Diaphragm seals of flange design For gage, absolute, differential pressure and flow with flexible capillary ■ Dimensional drawings D 2 ) n xd 2 _ ³1 R ø d4 ø k ø d4 ø k 50 n xd 2 b f dM dM (5. 9 D b f R ³1 50 R ³ (5. 150 9) (5. 9) + Connection to EN 1092-1 Nom. diam. Nom. press.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Diaphragm seals of flange design For gage and absolute pressure, directly fitted on transmitter ■ Overview 2 Diaphragm seals of flange design, directly fitted on a pressure transmitter for pressure ■ Technical specifications Diaphragm seals (flange design) for pressure and absolute pressure, directly fitted on a transmitter Capillary • Length Max. 10 m (32.8 ft), longer lengths on request PN 40, PN 100 • Internal diameter 2 mm (0.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Diaphragm seals of flange design For gage and absolute pressure, directly fitted on transmitter Selection and Ordering data Order No. Ord.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Diaphragm seals of flange design For gage and absolute pressure, directly fitted on transmitter ■ Dimensional drawings Connection to EN 1092-1 Nom. diam. Nom. press. b D DN 50 PN 40 mm mm mm 20 165 18 PN 100 28 195 26 102 Vertical process connection 193 (7.6) 2 83 (3.27) DN 80 f L b n x d2 f 193 (7.6) b mm mm n mm mm mm 102 48.3 59 2 125 4 48.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Diaphragm seals of flange design For differential pressure, fixed connection and with capillary ■ Overview 2 Diaphragm seals of flange design for pressure transmitters for differential pressure, fixed connection and with flexible capillary ■ Technical specifications Diaphragm seals of flange design for pressure transmitters for differential pressure, fixed connection and with flexible capillary Nominal diameter Nominal pressure • DN 80 P
© Siemens AG 2008 SITRANS P measuring instruments for pressure Diaphragm seals of flange design For differential pressure, fixed connection and with capillary Order No. Ord.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Diaphragm seals of flange design For differential pressure, fixed connection and with capillary ■ Dimensional drawings Connection to EN 1092-1 D dM Nom. diam. Nom. b press. mm mm mm mm mm mm mm DN 80 PN 40 24 200 18 138 76 89 2 160 8 DN 100 PN 16 20 200 18 158 94 89 2 180 8 PN 40 24 235 22 162 94 89 2 190 8 b mm f ø d4 ø k R _> 15 0( 5.9 ) n xd 2 D lb/ sq.in.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Quick-release diaphragm seals For gage, absolute and differential pressure ■ Overview - DN 65 PN 25 - DN 80 PN 25 • Clamp connection 2 - 1½ inch PN 40 - 2 inch PN 40 - 2½ inch PN 25 - 3 inch PN 25 For differential pressure and flow • To DIN 11851 with slotted union nut - DN 50 PN 25 - DN 65 PN 25 - DN 80 PN 25 • To DIN 11851 with threaded socket Quick-release diaphragm seals, to DIN 11851 with slotted union nut - DN 50
© Siemens AG 2008 SITRANS P measuring instruments for pressure Quick-release diaphragm seals For gage, absolute and differential pressure Order No. Ord. code Selection and Ordering data Quick-release diaphragm seal D) 7 M F 4 9 4 0 - for SITRANS P pressure transmitters for pressure 7MF403 7 and 7MF4237 together with Order code "V01" (vacuum-proof design) and 7MF80271); must be ordered separately Filling liquid: Food oil (FDA listed) Material: Stainless steel, mat. No. 1.4435 Nom. diam. Nom. press.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Quick-release diaphragm seals For gage, absolute and differential pressure ■ Dimensional drawings H H H 146 (5.75) 2 dM dM DN D D G1 dM dM DN dM D D dM G1 DN DN Mounted directly on SITRANS P transmitter for pressure R ³ 15 0( 5.9 ) G1 DN DN G1 Mounted on SITRANS P transmitter for pressure or differential pressure and flow Clamp connection (left) DN (1½ inch) (2 inch) (2½ inch) (3 inch) Ø dM 32 (1.26) 40 (1.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Miniature diaphragm seal For gage and absolute pressure ■ Overview ■ Technical specifications Miniature diaphragm seals Span with • G1B • G1½B • G2B > 6 bar (> 87 psi) > 2 bar (> 29 psi) > 600 mbar (> 8.7 psi) Filling liquid Silicone oil M5 or food oil (FDA listed) Material • Main body • Diaphragm Stainl. steel mat No. 1.4404/ 316L Stainl. steel mat No. 1.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Flushing rings For diaphragm seals ■ Overview ■ Technical specifications Flushing ring for remote seals of sandwich and flange design 2 Nominal diameter Nominal pressure • DN 50 PN 16 ... PN 100 • DN 80 PN 16 ... PN 100 • DN 100 PN 16 ... PN 100 • DN 125 PN 16 ... PN 100 • 2 inch Class 150 ... class 600 • 3 inch Class 150 ... class 600 • 4 inch Class 150 ... class 600 • 5 inch Class 150 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Flushing rings For diaphragm seals Selection and Ordering data Order No. Ord. code D) 7 M F 4 9 2 5 - 1 7777 Nom. diam. • DN 50 • DN 80 • DN 100 • DN 125 Nom. press. PN 16 ... PN 100 PN 16 ... PN 100 PN 16 ... PN 100 PN 16 ... PN 100 A B C D • 2 inch • 3 inch • 4 inch • 5 inch Class 150 ... 600 Class 150 ... 600 Class 150 ... 600 Class 150 ... 600 G H J K Other version Add Order code and plain text: Nominal diameter: ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Clamp-on seals of flange design For gage pressure, differential pressure and flow ■ Overview ■ Technical specifications Clamp-on seals for flange-mounting 2 Nominal diameter Nominal pressure • DN 25 PN 6 ... PN 100 • DN 40 PN 6 ... PN 100 • DN 50 PN 6 ... PN 100 • DN 80 PN 6 ... PN 100 • DN 100 PN 6 ... PN 100 • 1 inch Class 150 ... class 2500 • 1½ inch Class 150 ... class 2500 • 2 inch Class 150 ...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Clamp-on seals of flange design For gage pressure, differential pressure and flow Selection and Ordering data Order No. Ord. code Order code Selection and Ordering data Clamp-on seal for flange-mounting for SITRANS P pressure transmitters Further designs Please add "-Z" to Order No. and specify Order code.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Clamp-on seals of flange design For gage pressure, differential pressure and flow ■ Dimensional drawings Connection to EN 1092-1 2 146 (5.75) DN PN D Mb L mm bar mm mm mm H mm 25 6 ... 100 63 28,5 60 78.5 40 6 ... 100 85 43 60 89.5 50 6 ... 100 95 54.5 60 92.5 80 6 ... 100 130 82.5 60 112 100 6 ... 100 150 107 60 122 D Mb L H mm mm mm mm (inch) (inch) (inch) (inch) 63 28.5 60 78.5 (2.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Quick-release clamp-on seals For pressure and absolute pressure ■ Overview ■ Function The measured pressure is transferred from the diaphragm, mounted on the inner circumference of the clamp-on seal, to the filling liquid and then passes through the capillary to the measuring chamber of the pressure transmitter.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Quick-release clamp-on seals For pressure and absolute pressure Selection and Ordering data Quick-release clamp-on seal 2 for SITRANS P pressure transmitters for pressure 7MF403 7 and 7MF4237 together with Order code "V01" (vacuum-proof design) and 7MF80271); must be ordered separately Filling liquid: Food oil (FDA listed) Material: Stainless steel 316L Nom. diam. Nom. press.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Quick-release clamp-on seals For pressure and absolute pressure ■ Dimensional drawings H D D1 d D2 1 G D1 D2 DN H 146 (5.9) 2 L L L D D1 d D2 1 G 1 D D2 DN R ³ 15 0( 5.9 ) Mounted directly on SITRANS P transmitter for pressure L Mounted on SITRANS P transmitter for pressure or differential pressure and flow Connection to DIN 11851 with screw necks Clamp connection for pipes to BS 4825/3 and o.D.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Remote seals Measuring setups ■ Overview Possible combinations of pressure transmitters and remote seals This section shows examples of typical measuring setups for using SITRANS P pressure transmitters with and without remote seals.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Remote seals Measuring setups with remote seals ■ Dimensional drawings Types of installation for pressure and level measurements (open vessels) Installation type A Installation type B Installation type A 2 Start-of-scale: pMA = rFL · g · HU - rOil · g · H1 Full-scale: DH DH HO Installation type B HO Start-of-scale: pMA = rFL · g · HU + rOil · g · H1 Full-scale: HU HU H1 pME = rFL · g · HO - rOil · g · H1 H1 Legend pMA pME rFL rO
© Siemens AG 2008 SITRANS P measuring instruments for pressure Remote seals Measuring setups with remote seals Types of installation for level measurements (closed vessels) Installation type E Installation type E Start-of-scale: pMA = rFL · g · HU - rOil · g · HV Full-scale: 2 Legend pMA pME rFL rOil DH HV g HU HO HV HO HU Installation type G Installation type H pME = rFL · g · HO - rOil · g · HV Start-of-scale value to be set Full-scale value to be set Density of medium in vessel Density of fill
© Siemens AG 2008 SITRANS P measuring instruments for pressure Remote seals Measuring setups without remote seals ■ Overview Notes • For the separation layer measurement, the separation layer has to be positioned between the two spigots. Also you must make sure that the level in the container is always above the top spigot. • When measuring density, make sure that the level of the medium remains constant.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Remote seals Measuring setups without remote seals Measuring setups for closed containers Level measurement, Version 1 Static pressure 2 Start-of-scale: DpMA = r · g · HU Gas-filled negative pressure line Maximum level Full-scale value Start-of-scale value Full-scale: DpME = r · g · HO Legend DpMA DpME r g HU HO Start-of-scale value to be set Full-scale value to be set Density of medium in vessel Local acceleration due to gravity Star
s © Siemens AG 2008 Questionnaire Checking of transmitter/remote seal combinations * Customer:___________________ * Plant: ___________________ * Ordering code:________________ * Ordering department: _____ * Transmitter Order No.: 7MF 4 Tag. No.: _____________________ Item No.: _____________________ Person responsible: ___________ Phone: _____________________ –1 –1 2 Order No. of transmitter known? Yes No * Order No.
© Siemens AG 2008 s Questionnaire for hydrostatic level measurements Order date: _______________________________________________ Processing date: __________________________________________ 2 Ordering code (customer): _________________________________ Ordering code (supplier): __________________________________ Customer reference:_______________________________________ Measuring point: __________________________________________ Position:__________________________________________________ Dimensions: ____
© Siemens AG 2008 s Questionnaire (suitable for US market) Checking of transmitter/remote seal combinations * Customer:___________________ * Plant: ___________________ * Ordering code:________________ * Ordering department: _____ * Transmitter Order No.: 7MF 4 Tag. No.: _____________________ Item No.: _____________________ Person responsible: ___________ Phone: _____________________ –1 –1 2 Order No. of transmitter known? Yes No * Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fittings Technical description ■ Overview Classification according to pressure equipment directive (DGRL 97/23/EC): All 21-off fittings can be secured onto walls, racks (72 mm grid) and vertical and horizontal pipes. 2 This offers the advantage when assembling a plant that the shutoff fittings can be secured first and the lines for the medium and differential pressure connected to them.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fittings Selection aid ■ Selection of available shut-off valves Transmitters Shut-off valves for general applications Page Shut-off valves for special applications Relative and absolute presShut-off valves/double shutsure transmitters with process off valves to DIN 16270, connection G½" male thread DIN 16271 and DIN 16272 e.g. • SITRANS P, Z series 7MF1564-.....-.A.. • SITRANS P300 7MF802.-...0.-.... • SITRANS P DS III series 7MF403.-...0.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fittings Selection aid 2 Transmitters Shut-off valves for general applications Differential pressure transmitter with process connection to IEC 61518 e.g. • SITRANS P DS III series 7MF443.-... and 7MF453.-... For 3/5-spindle valve manifold 2/201 DN 5 7MF9411-5B. and 7MF9411-5C. PN 100 multiway cocks 7MF9004-...
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for gage and absolute pressure transmitters Shut-off valves to DIN 16270, DIN 16271 and DIN 16272 ■ Overview Order No. Selection and Ordering data Shut-off valves, form B, DIN 16270 without test collar, pipe union with ferrule 12 S DIN EN ISO 8484-1, without certificate Material Valve housing Maximum permissible working pressure P250GH (mat. No. 1.0460) 400 bar 7MF9401-8AB X 6 CrNiMoTi 17 12 2 400 bar (mat.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for gage and absolute pressure transmitters Shut-off valves to DIN 16270, DIN 16271 and DIN 16272 ■ Characteristic curves Operating pressure Steel or stainless steel version 400 bar at 120 °C 350 bar at 200 °C 300 200 Brass version 250 bar at 120 °C 200 bar at 200 °C 100 0 0 100 200 300 400 °C Operating temperature Permissible operating pressure as a function of the permissible operating temperature ■ Dimens
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for gage and absolute pressure transmitters Angle adapter ■ Overview ■ Dimensional drawings * ® É ၂ É * ® 2 P300 pressure transmitter with shut-off valve and angle adapter Angle adapter, dimensions in mm The angle adapter enables pressure transmitters with top displays to be read from the front. Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for gage and absolute pressure transmitters Double shut-off valves A B C 300 18 33 63.5 100.5 17 diam. 31.75 6.4 approx.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for gage and absolute pressure transmitters Accessories for shut-off valves / double shut-off valves ■ Overview ■ Overview The mounting set is suitable for the double shut-off valves 7MF9011-4.A and for wall, rack and pipe mounting. Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for gage and absolute pressure transmitters 2-way valve manifolds DN 5 ■ Overview ■ Function The characteristic of the pressure measuring instrument or pressure transmitter can be tested through the test connection. A pressure gage for local display can be connected to the test connection. 2 The two-way valve manifold DN 5 can be used in addition to shut off and vent the line with the measured medium.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for gage and absolute pressure transmitters 2-way valve manifolds DN 5 Selection and Ordering data Order No. Selection and Ordering data 2-way valve manifold DN 5 for flanging to pressure transmitters, max. working pressure 420 bar, weight 1.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for gage and absolute pressure transmitters 2-way valve manifolds DN 5 ■ Characteristic curves 300 200 100 0 0 100 200 300 400 °C Operating temperature 2-way valve manifold DN 5, permissible working pressure as a function of the permissible working temperature ■ Dimensional drawings 19,5 2 51 1 59 C 41 Ø12 B 7 M10, 12 2-way valve manifold DN 5 (7MF9401-2J/-2K) for connection to pressure gages and pressure
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 2-, 3- and 5-spindle valve manifolds DN 5 ■ Overview Selection and Ordering data Order No. Valve manifolds DN 5 for liquids and gases, for flanging to pressure transmitters for absolute and differential pressure, max.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 2-, 3- and 5-spindle valve manifolds DN 5 Order code Order No. Further designs1) Accessory set for 2-, 3- and 5-spindle valve manifolds Please add "-Z" to Order No. and specify Order code. 2 for valve manifolds 7MF9411-5B. and -5C. 4x screws M10x45 to DIN EN 24014; chromized steel 4x washers Ø 10.5 mm to DIN 125; 2x flat gaskets made of PTFE, max.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 2-, 3- and 5-spindle valve manifolds DN 5 ■ Dimensional drawings ¼-18 NPT 55 63.5 41.3 ½-14 NPT B C 2 A M10 Ø12 31.75 54 68 182 A Process connection: ½-14 NPT B Transmitter connection: Flange connection to EN 61 518, form B C Vent / test connection: ¼-18 NPT Valve design: external spindle thread Mounting plate 7MF9006-6..
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters Multiway cocks PN 100 ■ Overview ■ Technical specifications Multiway cocks PN 100 2 Measured medium Water, non-aggressive liquids and gases Aggressive liquids, gases and vapors Material P250GH, mat. No.: 1.0460 X 6 CrNiMoTi 17 12 2, mat. No. 1.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters Multiway cocks PN 100 ■ Accessories ■ Dimensional drawings Accessory set for multiway cock PN 100 • L31: 4 screws 7/16-20 UNF x 1 inch, 2 flat gaskets • L11: 4 screws M10x25 to DIN EN 24017, 4 washers, 2 flat gaskets • L15 (suitable for oxygen): 4 screws M10x25 to DIN EN 24017, 4 washers, 2 flat gaskets 2 Washers Ø 10.5 to DIN 125 Flat gaskets made of PTFE, max.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 3-way and 5-way valve manifolds DN 5 ■ Overview ■ Function • Shutting off the differential pressure lines • Checking the pressure transmitter zero • In addition, the five-way valve manifold permits blowing out of the differential pressure lines. 2 The three-spindle and five-spindle valve manifolds DN 5 (7MF9410-1../-3..
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 3-way and 5-way valve manifolds DN 5 Selection and Ordering data Order code Order No. Further designs1) Please add "-Z" to Order No. and specify Order code. Accessory set to EN (required for flanging, weight 0.2 kg) 7/ 4x screws 16-20 UNF x 21/8 inch to ASME B18.2; chromized steel 2x flat gaskets made of PTFE, max.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 3-way and 5-way valve manifolds DN 5 ■ Characteristic curves 300 200 100 0 0 100 200 300 400 °C Operating temperature Permissible operating pressure as a function of the permissible operating temperature Mounting plate 7MF9006-6..
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 3-way valve manifold DN 8 ■ Overview ■ Function The 3-way valve manifold DN 8 performs two functions as standard: • Shutting off the differential pressure lines • Checking the pressure transmitter zero All versions are also available with a test connection, to which a test device for checking the pressure transmitter characteristic can be connected.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 3-way valve manifold DN 8 Order code Order No. Further designs1) Please add "-Z" to Order No. and specify Order code. 2 Accessory set to EN (required for flanging, weight 0.2 kg) 4x screws 7/16-20 UNF x 21/8 inch to ASME B18.2; chromized steel 2x flat gaskets made of PTFE, max. permissible 420 bar, 80 °C B31 F) 7MF9010-5CC 4x screws 7/16-20 UNF x 21/8 inch to ASME B18.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 3-way valve manifold DN 8 ■ Dimensional drawings ■ Schematics 48 A C Ø12 D 2 D B 41,3 A 38 25 A 31,5 48 7 B B Three-way valve manifold for liquids and gases 8 M10 A Process connection B Transmitter connection C Vent/test connection 54 72 108 147 160 195 22 3-way valve manifold DN 8, connections 41 A Process connection (e.g.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters Valve manifold combination DN 5/DN 8 ■ Overview ■ Function • Shutting off the differential pressure lines • Blowing out the differential pressure lines • Checking the pressure transmitter zero 2 As an option it is possible to order a version with a test connection, to which a test device for checking the transmitter characteristic can be connected.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters Valve manifold combination DN 5/DN 8 ■ Accessories ■ Dimensional drawings Accessory set for valve manifold combination DN 5/DN 8 for flanging • B34: 4 screws 7/16-20 UNF x 21/8 inch to ASME B18.2.1, 2 O-rings (FPM 90) • B16: 4 screws M10x55 to DIN EN 24014, 4 washers, 2 O-rings (FPM 90) 2 Washers Ø 10.5 to DIN 125 O-ring to DIN 3771, 20 x 2.65 - S – FPM90, max.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters Valve manifold combination DN 8 ■ Overview ■ Function • Shutting off the differential pressure lines • Blowing out the differential pressure lines • Checking the pressure transmitter zero 2 As an option it is possible to order a version with a test connection, to which a test device for checking the pressure transmitter characteristic can be connected.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters Valve manifold combination DN 8 Selection and Ordering data Order No. Valve manifold combination DN 8 for vapors for flanging to pressure transmitters for differential pressure, with mounting plate, max.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 2-, 3- and 5-spindle valve manifolds for installing in protective boxes ■ Overview 2 Selection and Ordering data Order No. Valve manifolds DN 5 for mounting in protective boxes for liquids and gases for flanging to pressure transmitters for absolute and differential pressure Material: stainless steel, mat. No: 1.4404/316L max.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 2-, 3- and 5-spindle valve manifolds for installing in protective boxes Order code Order No. Further designs1) Accessory set for 2-, 3- and 5-spindle valve manifolds (Connection between manifold and transmitter) Please add "-Z" to Order No. and specify Order code. Accessory set to DIN (connection between valve manifold and pressure transmitter) for valve manifold 7MF9412–1C.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 2-, 3- and 5-spindle valve manifolds for installing in protective boxes ■ Dimensional drawings 12 54 G½ 37 2 115 41.3 B 80 63.5 M8 103 ½NPT 80 A M8 ¼-18 NPT C A 10 M8 ½-14 NPT 38 88.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 12 2-, 3- and 5-spindle valve manifolds for installing in protective boxes 72 100 2.5 72 55 4x Ø11 R5 2 8 72 8 0 R2 72 M8 12 +1 R5 R10 R2.5 60,3 100 155 17 9 R 19 Ø9 26.6 31.6 57.6 77.1 38 76 31.5 Ø16 Ø26 24 6.5 26 4.5 19 180 180 110 6.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 3- and 5-spindle valve manifolds for vertical angular differential pressure lines ■ Overview 2 Selection and Ordering data Order No. Valve manifolds for vertical differential pressure lines for liquids and gases for flanging to pressure transmitters for absolute and differential pressure Material: stainless steel, mat. No: 1.4404/316L max.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 3- and 5-spindle valve manifolds for vertical angular differential pressure lines Accessory set (connection between manifold and transmitter) • K36: 4 screws 7/16-20 UNF x 1¾ inch to ASME B18.2.1, 2 flat gaskets • K16: 4 screws M10x45 to DIN EN 24014, 4 washers, 2 flat gaskets Washers Ø 10.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters 100 72 5 3- and 5-spindle valve manifolds for vertical angular differential pressure lines 8 2 72 76 16 19 50 9 38 92 26 Mounting bracket (7MF9006-6QA)/(M19) for 3-spindle and 5-spindle valve manifolds, dimensions in mm ■ Schematics A A A Process connection B Transmitter connection B B 3-spindle valve manifold for vertical differential pressure lines, connection
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters Low-pressure multiway cock ■ Overview Order No. Selection and Ordering data Low-pressure multiway cock for liquids and gases, for flanging to pressure transmitters, max. working pressure 25 bar, max. working temperature 60 °C (up to 80 °C for a short time), weight 1.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fitttings - Shut-off valves for differential pressure transmitters Low-pressure multiway cock ■ Options Low-pressure multiway cock, permissible operating pressure as a function of the permissible operating temperature ■ Dimensional drawings Low-pressure multiway cock 7MF9004-4CA/-4DA for direct flanging to pressure transmitters for differential pressure, dimensions in mm 40 6 90 19 70 2 ■ Characteristic curves Test connections • 2 se
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fittings - Accessories Oval flange ■ Overview Order No. Selection and Ordering data Oval flange with female thread ½-14 NPT, max. working pressure 420 bar, flange connection to DIN EN 61518, form A 2 Material P250GH, mat. No.: 1.0460 7MF9408-2CE X 2 CrNiMo 17 13 2, mat. No. 1.4404/316L 7MF9408-2CL Order code Order No. 2x screws 7/16-20 UNF x 1½ inch to ASME B 18.2.3; chromized steel 1x flat gasket made of PTFE, max.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fittings - Accessories Adapters, connection glands ■ Overview ■ Dimensional drawings Adapters enable e.g. a transition from medium connections with NPT thread to shut-off valves to DIN 16270 ... 16272 or pipes in conjunction with a connection gland (e.g. 7MF9008). 2 ■ Design The connection pieces are made of X 6 CrNiMoTi 17 12 2, mat. No. 1.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fittings - Accessories Connection parts G 1/2 ■ Dimensional drawings Selection and Ordering data CW 614N Union nut 9 SMn 28 k Nipple: RSt 37-2 16 M20 x 1,5 30 16 Ø7 SW 27 M56340-A0001 M56340-A0002 1.0715 Ø 14 Nipple connection (M20 x 1.5) M56340-A0008, dimensions in mm 1.0037 Union nut X 8 CrNiS 18 9 1.4305 Nipple: X 6 CrNiMoTi 17 12 2 1.4571/316Ti M56340-A0003 Nipple connection M20 x 1.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fittings - Accessories Water traps, Sealing rings to EN 837-1 ■ Overview ■ Overview Water traps protect pressure gages and shut-off fittings from heating up (e.g. by steam) by the water column produced by the water trap. The max. working temperature is 120 °C at 160 bar, 300 °C at 80 bar or 400 °C at 63 bar.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fittings - Accessories Pressure surge reducers ■ Overview The pressure surge reducer protects the pressure gage against damage, premature wear and tear and inaccurate/fluctuating indications. 2 ■ Application The pressure reducer is used when pulsations occur in the measured medium (e.g. in slow-running vapor engines, piston pumps and compressors), or if drastic fluctuations are likely to occur in the measured medium (e.g.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fittings - Accessories Primary shut-off valves ■ Overview Ø80 95 The shut-off valves are available in various materials and with various connections (see Selection and Ordering data) ■ Characteristic curves 500 Ø11,2 Ø11,2 bar 400 G½ 22 300 Operating pressure C B 200 D E 90 22 G½ Shut-off valve 7MF9017–1A.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fittings - Accessories Primary shut-off valves ØAxb 7MF9017- 14 mm x 2.5 mm 1F. and 1G. 21.3 mm x 6.3 mm 1H. and 2H. 24 mm x 7.1 mm 1J., 1K. and 2J. 2 141 Ø120 Øa Ø14 b 2,5 5 65 4,5 130 Shut-off valves 7MF9017-, dimensions in mm Selection and Ordering data Primary shut-off valves, without certificate Max. Charac- Material Mat. No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fittings - Accessories Compensation vessels ■ Overview ■ Dimensional drawings The compensation vessels prevent the level difference which occurs with pressure changes in the pressure lines and which falsifies the measurement. 9 According to DIN 19211, the temperature in the compensation vessel must be assumed to be 50 K less than the steam temperature in the pipe when calculating the wall thicknesses.
© Siemens AG 2008 SITRANS P measuring instruments for pressure Fittings - Accessories Connection parts ■ Overview ■ Dimensional drawings Connection parts are available in the following versions: • Threaded flange pair G½ with stainless steel gasket • Nipple G½ form V to DIN 19207 • Union nut G½ made of C 35 to DIN 16284 • Gasket B½ (grooved) to DIN 19207 2 All connection parts are also available grease-free for oxygen. Selection and Ordering data Order No.
© Siemens AG 2008 SITRANS P measuring instruments for pressure 2 2/234 Siemens FI 01 · 2009