Operating instructions
205
Children
For the 2
nd
reconstruction the Autoload into MPRthick
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MPRthick, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset, adapt the parameters to your needs and link
them to the series.
Contrast medium IV injection
Start delay exam dependent
Flow rate dependent upon needle size/Access site
Total amount 1 – 2 ml per kg of body weight
* The conversion factor for an 8-week-old, and a scan
range of 30 mm was used.
SinusOrbi 2
nd
reconstr.
kV 120
Effective mAs 60
Rotation time 0.5 sec.
Slice collimation 0.75 mm
Slice width 3.0 mm 1.0 mm
Feed/Rotation 5.4 mm
Increment 3.0 mm 0.7 mm
Kernel H60f H60f
CTDI
Vol
12.6 mGy
Effective dose (mSv) Male: 0.5* Male: 0.5*
Female: 0.4* Female: 0.4*