Children
Hints
• Children, who are more than 6 years old, should be
scanned with an adult protocol.
• The second recon job is defined with kernel H60s
and with an overlap for visualizing bone structures
with MPR.
• If the study is focused on bone structure only, the
mAs can be reduced to 20 for all age groups.
• In order to optimize image quality versus radiation
dose, scans are provided within a maximum scan field
of 300 mm with respect to the iso-center. No recon
job with a field of view exceeding those limits will be
possible. Therefore, patient positioning has to be
performed accurately to ensure a centered location
of the skull.
• To work without CARE Dose 4D use for children
< 3 years 40 mAs