Operating instructions
208
Children
NeckRoutine
Indications:
Spiral mode for routine neck studies, e.g. tumors,
lymphoma, abscesses, etc.
A typical range of 17 cm covered in 8 sec.
Contrast medium IV injection
Start delay exam dependent
Flow rate dependent upon needle size/Access site
Total amount 1 – 2 ml per kg of body weight
* The conversion factor for an 8-week-old, and a scan
range of 80 mm was used.
NeckRoutine 2
nd
reconstr.
kV 120
Effective mAs 60
Rotation time 0.75 sec.
Slice collimation 1.5 mm
Slice width 6.0 mm 2.0 mm
Feed/Rotation 18.0 mm
Increment 6.0 mm 1.5 mm
Kernel B30s B60s
CTDI
Vol
4.2 mGy
Effective dose (mSv) Male: 0.7* Male: 0.7*
Female: 0.8* Female: 0.8*