Operating instructions

209
Children
For the 2
nd
reconstruction the Autoload into MPRthick
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MPRthick, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset, adapt the parameters to your needs and link
them to the series.
Hints
• If necessary, scan down to the aortic arch or
mediastinum to include the entire lesion.
• Cooperative children can be instructed to hold their
breath during the acquisition.
• Children, who are more than 6 years old should be
scanned with an adult protocol.
• To work without CARE Dose 4D use for children
< 3 years 40 mAs
3-6 years 60 mAs
6-12 years 120 mAs