Operating instructions
312
Pelvis
SI_Joints
Indications:
Spiral mode for the sacroiliac joints, e.g. evaluation
of joint cavity, masses, trauma, dysplasia, necrosis,
congruence evaluations, orthopedic indications etc.
A typical range of 15 cm will be covered in 5.0 sec.
SI_Joints 2
nd
reconstr.
kV 120
Effective mAs 160
Slice collimation 0.75 mm
Slice width 3.0 mm 1.0 mm
Feed/Rotation 9.0 mm
Rotation time 0.5 sec.
Kernel B60f B60f
Increment 3.0 mm 0.7 mm
CTDI
Vol
12.48 mGy
Effective dose Male: 2.4 mSv
Female: 2.1 mSv
For the 2
nd
reconstruction the Autoload into MPRthick
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MPRthick, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset, adapt the parameters to your needs and link
them to the series.