Operating instructions

HeartView CT
143
Preview Series
A Heart View preview series contains images at the
same slice position and of different phases of the car-
diac cycle (Phase Start). The phase interval and the
number of images can be configured in the HeartView
configuration. The slice position is determined by
selecting the appropriate image in the right segment.
The preview series is used to determine the Phase Start
for ED and ES.
For an overview, reconstruct a series at an arbitrary
heart phase. A default value of 60% relative gating
can be used as an initial set-up for the overview
series.
Select an appropriate slice position (e.g. middle part
of RCA) for the preview series in the 2
nd
segment.
Switch to the Trigger tabcard.
•Select Phase Start 0 % to reconstruct images in %
steps.