Operating instructions

HeartView CT
153
Synthetic Trigger
By default, the Synthetic Trigger (ECG-triggered scan-
ning) or Synthetic Sync (ECG-gated scanning) is acti-
vated for all predefined cardiac scan protocols. It is rec-
ommended to always keep it activated for
examinations with contrast medium.
In case of ECG signal loss during the acquisition, this
will ensure the continuation of the triggered scans or
allows an ECG to be simulated for retrospective gating.
If it is deactivated, the scanning will be aborted in case
of ECG signal loss during the acquisition.