Operating instructions
156
HeartView CT
Double-Oblique Images
Reconstruction of double-oblique images are used for
post-processing.
If you want to use short axis images of the heart, there
are different possibilities to create them:
• During scanning
You can reconstruct short axis spiral oblique (SPO)
images in different heart phases directly during the
examination.
• After scanning
1.If you have saved the raw data of the Cardio scan:
You can reconstruct short axis spiral oblique (SPO)
images in different heart phases by loading the raw
data in the Examination card of your Navigator, or
you can also load the raw data in the reconstruction
card of your Wizard (optional) console.
2.After reconstruction of different axial heart phases,
you can reconstruct short axis MPR images on the 3D
card.