Operating instructions

162
HeartView CT
3.Reconstruction of irregular heart phases, e.g. for the
best display of the coronary arteries:
LAO:
Select Multiphase settings: Manual and choose %
values.
Enter "50" "52" "55" "58" "60" to find out the best dis-
play of the LAO view.
As result you will get 5 series with the heart phases:
50%, 52%, 55%, 58%, 60%.
If you want to create your own multiphase recon job in
a cardio scan protocol:
Define either an auto or a manual setting, than select
the Multiphase settings – Protocol. Press button
Apply, close the window with OK. Then you can save
the scan protocol as usual and the selected recon job
will always be reconstructed using a multiphase.
Hints:
To get a constant image quality for the Multiphase
reconstruction we recommend turning ECG pulsing
OFF.
If you are not sure which heart phases are the best
one for the Multiphase reconstruction, perform a
preview series first.
•Press Recon button for each 3D reconstruction.