Operating instructions
166
HeartView CT
Auto Load in 3D and Postprocessing
Presets
You can activate the Auto load in 3D function on the
Examination card/Auto Tasking and link it to a recon
job, e.g. the 2
nd
recon job with thinner slice width in
some of the examination protocols. If the post-pro-
cessing type is chosen from the pull down menu, the
reconstructed images will be loaded automatically into
the 3D card on the Navigator with the corresponding
postprocessing type.
On the 3D card you have the ability to create Range
Parallel and Radial protocols for Multi-Planar-Recon-
struction (MPR) and Thin Maximum-Intensity-Projec-
tion (MIP Thin), which can be linked to a special series.
For example, if you always do sagittal Multiplanar
reconstructions for a Spine examination, once you
load a Spine examination into the 3D card, select the
image type (MPR, MIP Thin), select the orientation and
open the parallel ranges function. Adapt the range
settings (image thickness, distance between the
images etc.) and hit the link button. From that point
on, you have a predefined postprocessing protocol,
linked to the series description of a Spine examination.