Operating instructions

syngo Calcium Scoring
171
* depends on heart rate
CaSc
kV 120
Effective mAs/
Quality ref. mAs
150
Rotation Time 0.42 sec.
Acquisition 16 x 1.5 mm
Slice collimation 1.5 mm
Slice width 3.0 mm
Feed/Rotation 6.7 mm
Pitch Factor 0.28
Increment 1.5 mm
Kernel B35f
Temp. resolution* up to 105 ms
CTDI
Vol
10.5 mGy
Effective dose Male: 3.47 mSv
Female: 4.36 mSv