Operating instructions

178
syngo Calcium Scoring
Prerequisites
To achieve standardization of the evaluation results
the image material must meet the following require-
ments:
Only images of one patient must be loaded for
syngo Calcium Scoring evaluation.
The scans must have been acquired without gantry
tilt (gantry tilt = 0).
Only original image material must be used that has
neither been edited with 2D tools in the Viewing
task card e.g., Zoom&Pan nor with 3D tools in 3D
e.g., curved mode.
Only image data sets with the same slice orientation
must be used.
Only contiguously scanned and reconstructed
sequence or spiral data sets with identical slice dis-
tances and identical thickness must be evaluated.
The slices must have been scanned at different table
positions.
The data capacity of the volume must not exceed
1024 images with a matrix size of 512 x 512 pixels.
Only studies without contrast medium must be
used.