Operating instructions

364
Abdomen
AbdomenCombi
Indications:
Combination of thin slice and routine abdominal stud-
ies with one spiral scan.
A range of 20 cm for liver, pancreas or kidneys will be
covered in 12.11 sec.
AbdCombi 2
nd
reconstr.
kV 120
Effective mAs/
Quality ref. mAs
160
Rotation time 0.5 sec.
Acquisition 16 x 0.75 mm
Slice collimation 0.75 mm
Slice width 5.0 mm 1.0 mm
Feed/Rotation 9.0 mm
Pitch Factor 0.75
Increment 5.0 mm 0.7 mm
Kernel B30f B20f
CTDI
Vol
12.5 mGy
Effective dose Male: 4.99 mSv
Female: 6.36 mSv
Contrast medium IV injection
Start delay 50 – 65 sec.
Flow rate 4.0 ml/sec.
Total amount 100 ml