Operating instructions

372
Abdomen
AbdSeq
Indications:
This protocol is created for measurement with sequen-
tial mode in the region of the abdomen.
A scan range is predefined with 20.5 cm.
AbdSeq
kV 120
Effective mAs/
Quality ref. mAs
140
Rotation time 0.75 sec.
Acquisition 2 x 5.0 mm
Slice collimation 5.0 mm
Slice width 5.0 mm
Feed/Scan 10.0 mm
Kernel B31s
CTDI
Vol
8.8 mGy
Effective dose Male: 3.16 mSv
Female: 3.96 mSv
Contrast medium IV injection
Start delay 50 – 65 sec.
Flow rate 4 ml/sec.
Total amount 100 ml