Vascular
ThorAngioRoutine/
ThorAngio042s
Indications:
Spiral mode for thoracal CT Angios, e.g. visualization
of tumors, metastases, lymphoma, lymph nodes, vas-
cular anomalies etc.
A range of 40 cm will be covered in 15.49/13.01 sec.
ThorAngio 2
nd
reconstr.
kV 120
Effective mAs/
Quality ref. mAs
120
Rotation time 0.5/0.42 sec.
Acquisition 16 x 0.75 mm
Slice collimation 0.75 mm
Slice width 5.0 mm 1.0 mm
Feed/Rotation 13.8 mm
Pitch Factor 1.15
Increment 5.0 mm 0.7 mm
Kernel B30f B20f
CTDI
Vol
9.4 mGy
Effective dose Male: 6.14 mSv
Female: 7.6 mSv