Operating instructions
Vascular
457
BodyAngioRoutine
Indications:
Spiral mode for abdominal CT Angio studies.
A typical study in a range of 40 cm will be covered in
15.49 sec.
BodyAngio 2
nd
reconstr.
kV
120
Effective mAs/
Quality ref. mAs
140
Rotation Time 0.5 sec.
Acquisition 16 x 0.75 mm
Slice collimation 0.75 mm
Slice width 5.0 mm 1.0 mm
Feed/Rotation 13.8 mm
Pitch Factor 1.15
Increment 5.0 mm 0.7 mm
Kernel B30f B20f
CTDI
Vol
10.9 mGy
Effective dose Male: 7.16 mSv
Female: 8.87 mSv