Operating instructions
460
Vascular
A typical study of the whole aorta including its bran-
chiocephalic trunk and iliac arteries in a range of 40 cm
will be covered in 13.01 sec.
Hints
• CARE Bolus may be used to optimize bolus timing.
• Set the ROI for monitoring scan in the aorta abdom-
inalis with triggering threshold of 120 HU, or use
manual triggering.
• Do not administer oral contrast medium, as this
impairs the editing of MIP/SSD/VRT images.
• Use water as oral contrast.
• The length of the CM spiral in the topogram (via the
table position) can be planned exactly by means of
pre-contrast images.
BodyAngio
Fast
2
nd
reconstr.
kV
120
Effective mAs/
Quality ref. mAs
140
Rotation Time 0.42 sec.
Acquisition 16 x 0.75 mm
Slice collimation 0.75 mm
Slice width 5.0 mm 1.0 mm
Feed/Rotation 13.8 mm
Pitch Factor 1.15
Increment 5.0 mm 0.7 mm
Kernel B30f B20f
CTDI
Vol
10.9 mGy
Effective dose Male: 7.16 mSv
Female: 8.87 mSv
Contrast medium IV injection
Start delay 10 sec.
Flow rate 3.5 ml/sec.
Total amount 120 ml