Operating instructions

Vascular
467
AngioRunOff
Indications:
For CT Angio spiral studies of the extremities.
A range of 80 cm will be done in 18.54 sec.
AngioRunOff 2
nd
reconstr.
kV
120
Effective mAs/
Quality ref. mAs
140
Rotation Time 0.5 sec.
Acquisition 16 x 1.5 mm
Slice collimation 1.5 mm
Slice width 6.0 mm 2.0 mm
Feed/Rotation 22.8 mm
Pitch Factor 0.95
Increment 6.0 mm 1.4 mm
Kernel B30f B20f
CTDI
Vol
9.8 mGy
Effective dose Male: 6.43 mSv
Female: 6.19 mSv