Operating instructions

Specials
479
TraumaVol
Indications:
This is a one-range mode for fast screening for emer-
gency studies.
A scan range of 75 cm will be done in 14.59 sec.
Trauma 2
nd
recon. 3
rd
recon.
kV 120
Effective mAs/
Quality ref.
mAs
140
Rotation time 0.5 sec.
Acquisition 16 x
1.5 mm
Slice
collimation
1.5 mm
Slice width 7.0 mm 7.0 mm 7.0 mm
Feed/Rotation 27.6 mm
Pitch Factor 1.15
Increment 7.0 mm 7.0 mm 7.0 mm
Kernel B30f B30f B30f
CTDI
Vol
9.8 mGy
Effective dose Male: 12.86 mSv
Female: 14.41 mSv