Operating instructions
480
Specials
PolyTrauma
Indications:
Two combined ranges with two Topos are predefined,
head with neck and thorax with abdomen for emer-
gency studies.
A scan range of 12/14 cm will be done in 6.5/4.89 sec.
Head
kV 120
Effective mAs/
Quality ref. mAs
260
Rotation time 0.75 sec.
Acquisition 16 x 1.5 mm
Slice collimation 1.5 mm
Slice width 6.0 mm
Feed/Rotation 18.0 mm
Pitch Factor 0.75
Increment 6.0 mm
Kernel H31s
CTDl
Vol
49.4 mGy
Effective dose Male:
2.55 mSv