Operating instructions

Specials
505
Radiation exposure to personnel
During the procedure the physician is in the scan room
and close to the exposed scan plane.
Take special care to avoid excessive and unnecessary
radiation exposure.
Protection against primary x-ray exposure:
In the worst case, any body parts in the scan plane
may receive approximately the accumulated dose as
shown on the CTDI display.
Avoid being directly exposed to the x-ray beam.
Protection against stray radiation:
The x-ray beam is limited to the imaged scan vol-
ume, but a significant portion of the x-rays is scat-
tered and distributed in the scan room (stray radia-
tion).
Wear protective clothing to reduce exposure.
A table with measured data of this stray radiation is
included in the chapter on Safety in your SOMATOM
Operator Manual.