Operating instructions

Respiratory Gating
523
The respiratory gating hardware (AZ-733V) is com-
posed of a:
Respiratory Sensor (RS) to detect patient's abdomi-
nal motions (pressure changes) to get a patient's res-
piratory information,
Sensor Port (SP) to amplify and to transmit analog
signals from RS,
Wave Deck (WD) to receive the respiratory signal
from the said SP and to convert them to a digital sig-
nal to be sent to the host computer (PC) of the CT
system.
The respiration curve as well as scan and reconstruc-
tion parameters are displayed on the CT user interface
and embedded into the examination workflow.
The respiration curve will be displayed before and dur-
ing spiral acquisition and saved in the respiration file.
After scanning the user can select the respiration level
(in [%]) of inspiration or expiration for reconstruction.
A synthetic sync signal and a respiration curve editing
functionality are available.
Regarding an imaging device such as SOMATOM Sen-
sation Cardiac, the respiratory gating functionality
offers an increase of image resolution and reproduc-
ibility as well as a decrease of motion artifacts by
image reconstructions based on respiratory informa-
tion both needed for high-precision radiation therapy
planning and treatment as well.
The mentioned increase of RT accuracy results in a
decrease of RT side effects by minimizing excessive
irradiation dose to healthy tissue.