Operating instructions

Children
565
Hints
Children, who are more than 6 years old, should be
scanned with an adult protocol as the skull by this
time is fully grown.
The UHR mode has a maximum FoV of 300 mm. It is
mandatory to position the patient in the center of
the scan FoV.
When soft tissue is of interest, use kernel U30u for
image reconstruction.
In order to optimize image quality versus radiation
dose, scans are provided within a maximum scan
field of 300 mm with respect to the iso-center. No
recon job with a field of view exceeding those limits
will be possible. Therefore, patient positioning has to
be performed accurately to ensure a centered loca-
tion of the skull.
•To work without CARE Dose 4D use for children
< 3 years 40 mAs
3-6 years 60 mAs.
Contrast medium IV injection
Start delay exam dependent
Flow rate dependent upon needle size/Access
site
Total
amount
1 – 2 ml per kg of body weight