Children
AbdomenRoutine
Indications:
Spiral mode for routine studies in the region of abdo-
men and pelvis, e.g. tumors, lymphoma, abscesses,
post-traumatic changes, etc.
A typical range of 20 cm covered in 4.64 sec.
* The conversion factor for a 7-year-old child, and a
scan range of 200 mm was used.
AbdRoutine
kV 120
Effective mAs/
Quality ref. mAs
55
Rotation time 0.5 sec.
Acquisition 16 x 1.5 mm
Slice collimation 1.5 mm
Slice width 5.0 mm
Feed/Rotation 27.6 mm
Pitch Factor 1.15
Increment 5.0 mm
Kernel B30f
CTDI
Vol
3.9 mGy
Effective dose Male: 2.38 mSv*
Female: 2.9 mSv*