Dose Information
Note: The Image Quality Reference mAs corresponds
to the (eff.) mAs value that the radiologist would apply
for a reference patient without the use of CARE Dose
4D. The reference patient is defined as a typical adult,
70 kg to 80 kg or 155 to 180 lbs. (for adult protocols)
or as a typical child, 5 years, appr. 20kg or 45 lbs. (for
child protocols). After entering the Image Quality Ref-
erence mAs the CARE Dose type may be switched from
CARE Dose to CARE Dose 4D in the corresponding col-
umn. In the Scan Card the checkbox name will auto-
matically change from CARE Dose to CARE Dose 4D,
the Image Quality Reference mAs will be displayed and
in the Routine Card the comment CARE Dose 4D will
added to the (eff.) mAs value when CARE Dose 4D is
activated.
Note: When using CARE Dose 4D with child protocols,
make sure that all memos in the Chronicle with the cor-
relation between the patient’s weight and the mAs are
deleted! With CARE Dose 4D these memos loose their
meaning because the mAs are automatically adapted
to the patient size. If CARE Dose 4D is switched off
again, the mAs have to be manually adapted to the
patient size. Proposals for mAs settings may be taken
from the application guide.