Children
SpineThinSlice
Indications:
Spiral mode for the spine when Multi Planar Reforma-
tion (MPR) are intended, e.g. post-traumatic changes,
tumors, malformations, etc.
A typical range of 16 cm covered in 8.17 sec.
* The conversion factor for a 7-year-old child, and a
scan range of 160 mm was used.
SpineThinSlice 2
nd
reconstr.
kV 120
Effective mAs/
Quality ref. mAs
70
Rotation time 0.75 sec.
Acquisition 16 x 0.75 mm
Slice collimation 0.75 mm
Slice width 2.0 mm 1.0 mm
Feed/Rotation 18.0 mm
Pitch Factor 1.5
Increment 2.0 mm 0.7 mm
Kernel B30s B60s
CTDI
Vol
3.1 mGy
Effective dose Male: 1.51 mSv*
Female: 1.92 mSv*