Operating instructions
30
Aortic and Pulmonary Studies
This application can be used for high-resolution inter-
stitial lung studies with an ECG triggering technique,
or when imaging the aorta and pulmonary arteries
with contrast medium and ECG gating technique. The
following scan protocols are predefined:
• LungECGHires
– Sequential scanning protocol with ECG triggering
• PulmonaryECG
– Spiral scanning protocol with ECG gating
General Hints:
• The general purpose of these applications is to
reduce motion artifacts in the lungs due to the
cardiac pulsation.
• The LungECGHires protocol is recommended for
detection and localization of the lesions adjacent to
the heart or the interlobar fissures.
• The PulmonaryECG protocol is recommended for
aortic or pulmonary studies, e. g. aorta dissection or
pulmonary emboli.
HeartView CT