Operating instructions

31
HeartView CT
LungECGHires
Indications:
This is a sequential scanning protocol with an ECG
triggering technique for interstitial studies of the lungs,
especially for the lesions in the pericardial region.
LungECGHires
kV 120
Effective mAs 120
Slice collimation 0.75 mm
Slice width 0.75 mm
Feed/Scan 6.0 mm
Rotation time 0.75 sec.
Temporal resolution 375 ms
Kernel B80s
Image order cr-ca
CTDI
w
10.1 mGy
Effective dose Male: 1.1 mSv
Female: 1.4 mSv
Topogram:
AP, 512 mm.
From the apex of
the lung till the lung
base.