Operating instructions
64
Perfusion CT
How to do it
Scan protocols
There are two Scan protocols available:
BrainPerfCT, a non contrast study and a Dynamic
Serio with 12 mm collimation and 2 times a 12 mm
slice thickness.
BrainPerfSingle, a non contrast study and a Dynamic
Serio with 5.0 mm collimation and one 10 mm slice
thickness.
BrainPerfCT NonContrast DynSerio
kV 120 80
Effective mAs 320 140
Slice collimation 3.0 mm 12.0 mm
Slice width 9.0 mm 12.0 mm
Feed/Scan 18 mm 0 mm
Rotation time 1.0 sec. 0.5 sec.
Cycle time 3.0 sec. 0.75 sec.
Kernel H30s H30f
Image order ca-cr ca-cr
BrainPerfSingle NonContrast DynSerio
kV 120 80
Effective mAs 320 168
Slice collimation 3.0 mm 5.0 mm
Slice width 9.0 mm 10.0 mm
Feed/Scan 18 mm 0 mm
Rotation time 1.0 sec. 0.5 sec.
Cycle time 3.0 sec. 0.75 sec.
Kernel H30s H30f
Image order ca-cr ca-cr
Such a standard protocol may be slightly modified for
specific reasons; e. g. the start delay can be increased
by a few seconds for patients with very low cardiac
output.