Operating instructions

141
* CARE Bolus may be used to optimize the bolus timing.
Set the ROI for monitoring scan in the abdominal
aorta at the beginning of the scan range with trigger-
ing threshold of 120 HU, or use manual triggering.
The data acquired can also be used for image recon-
struction of soft tissue, done by the first reconstruction.
MIPthin: AngioRunOff coronal
Image thickness 3
Distance between images 3
Number of images 50
For the 2
nd
reconstruction the Autoload into the
MIPthin Range on the 3D Card is activated. The images
will be automatically loaded into 3D, MIPthin, and
coronal MIPthin Range will pop up.
If you are not satisfied with the Range preset, adapt the
parameters to your needs and link them to the series.
Extremities
Contrast medium IV injection
Start delay 25 30 sec.*
Flow rate 3.0 3.5 ml/sec.
Total amount 150 ml