Operating instructions
63
Bolus Tracking
• After the Topogram is performed, the predefined
spiral scanning range and the optimal monitoring
position will be shown.
• If you need to redefine the spiral scanning range,
you should also reposition the monitoring scan in
order to keep the shortest start delay time (2 s).
(The distance between the beginning of the spiral
scanning range and the monitoring scan will be
the same).
• Move the monitoring scan line towards the opti-
mal position and release the mouse button, it will
be snapped automatically. (Trick: if you move the
monitoring scan line away from the optimal posi-
tion the “snapping” mechanism will be inactive).
4. Place an ROI in the premonitoring scan on the
target area or vessel used for triggering with one
left mouse click. (The ROI is defined with double
circles – the outer circle is used for easy positioning,
and the inner circle is used for the actual evaluation).
You can also zoom the reference image for easier
positioning of the ROI.
5. Set the appropriate trigger threshold, and start
contrast injection and monitoring scans at the same
time.
During the monitoring scans, there will be simul-
taneous display of the relative enhancement of the
target ROI. When the predefined density is reached,
the spiral acquisition will be triggered automatically.
6. You can also initiate the spiral any time during the
monitoring phase manually – either by pressing the
START button or by left mouse clicking the START
radio button. If you do not want to use automatic
triggering, you can set your trigger threshold number
extremely high so that it will not trigger automa-
tically, and you can start the spiral when you desire.