Operating instructions
70
Interventional CT
5. Click “Load” and then “Cancel move”. Press the
“Start” button and 4 images will be displayed.
6. Press “Start” again, you’ll get another 4 images with
the same slice position.
BiopsyCombine
Indications: This is the biopsy mode with one com-
bined slices. It can be appended to any other scan
protocol, e. g. ThoraxRoutine for biopsy procedures
in the thorax.
Change the mAs setting accordingly before you load
the mode.
BiopsyCombine
kV 120
Effective mAs 120
Slice collimation 5.0 mm
Slice width 10 mm
Feed/Scan 0 mm
Rotation time 0.5 sec.
Kernel B30f
Application procedures:
1. Perform a spiral scan first to define a target slice.
2. Click “Same TP” under Table position in the routine
card, and move the table.
3. Turn on the light marker to localize the Entry point,
and then start the patient preparation.
4. Select “BiopsyCombine” mode under Special proto-
cols, and then click ”Append”.
5. Click “Load” and then “Cancel move”. Press the
“Start” button and one image will be displayed.
6. Press “Repeat” and “Start” again, you’ll get another
image with the same slice position.