Operating instructions

72
WholeBodyAngio
This scan protocol can be used for a whole Body Angio.
A range of 100 cm can be covered in 14.5 sec.
WholeBody 2
nd
Reconstr.
kV 120
Effective mAs 130
Slice collimation 1.5 mm
Slice width 6 mm 2 mm
Feed/Rotation 27.0 mm
Rotation time 0.5 sec.
Kernel B30f B20f
Increment 6 mm 1.5 mm
Image order cr-ca
CTDI
W
9.1 mGy
The reconstruction can be loaded in the 3D Card,
MPR. Select the Range function and a predefined range,
linked to the series, will pop up.
If you prefer to change this preset, adapt the parameters
to your needs and link them to the series.
MIPthick: WholeBodyAngio coronal
Image thickness 3
Distance between images 3
Number of images 60