Interval Trigger
Trigger when the times difference between the neighboring rising or falling edges meets
the time limit (<=, >=, [--,--], --][--).
To set interval trigger:
1. Press the
Setup
button on the front panel to enter the TRIGGER system function
menu.
2. Press the
Type
softkey; then use the
Universal Knob
to select
Interval
and push down
the knob to confirm.
3. Press the
Source
softkey; turn the
Universal Knob
to select CH1 or CH2 as the trigger
source.
4. Press the
Slope
softkey to select rising or falling edge.
5. Press the
Limit Range
softkey; turn the
Universal Knob
to select desired condition.
<= (less than a time value):
trigger when the positive or negative pulse time of
the input signal is lower than the specified time value.
>= (greater than a time value):
trigger when the positive or negative pulse time of
the input signal is greater than the specified time value.
[--,--] (within a range of time value):
trigger when the positive or negative pulse
time of the input signal is greater than the specified lower limit of time and lower
than the specified upper limit of time value.
--][-- (outside a range of time value):
trigger when the positive or negative pulse
time of the input signal is greater than the specified upper limit of time and lower
than the specified lower limit of time value.
6. Press the
Time Setting
softkey (
<=, >=, [--,--],--][--
), turn the
Universal Knob
to select
the desired value.