5 ℃. Make sure that the oscilloscope has been warmed up or operated for more than 30
minutes before the self-calibration.
Do the following steps to do self-calibration:
1. Disconnect all the input channels.
2. Press the
Utility
button on the front panel,and then press the
Do Self Cal
softkey, and
the oscilloscope will pop-out the message box shown as below:
Figure 109 Do Self Cal
3. Press the Single button on the front panel to perform the self-calibration program.
During the calibration, most of the keys are disabled.
4. When the self-calibration program is finished, it will pop-out the message “
press
Run/Stop key to exi
t”. Press the
Run/Stop
button on the front panel to exit the
calibration interface