User manual

SIGLENT
SDM3055 Digital Multimeter 95
Loading Errors (AC Voltage)
In the AC Voltage function, the input of SDM3055 appears as a 1MΩ
resistance in parallel with 100pF of capacitance. The test lead that you use to
connect signals to the Multimeter will also add additional capacitance and
loading. The approximate input resistances of the Multimeter at different
frequencies are listed in the following table.
Table 5- 2 Approximate Input Resistances at Different Frequencies
Input Frequency Input Resistance
100Hz 1MΩ
1kHz 850kΩ
10kHz 160kΩ
100kHz 16kΩ
For low frequencies:
Error%= ×100%
For high frequencies:
Error%=[ -1] ×100%
F: input frequency
R
s
: source resistance
C
m
: input capacitance (100pF) plus test lead capacitance