58 SDM3055 Digital Multimeter
Single Trigger
Single Trigger parameters that need to be set up include delay, samples
/trigger and VMC out.
Operating Steps:
1. Press , then select Trg Src】→【Singleor press on the
front panel directly to enable Single Trigger. See the diagram below.
Diagram 2- 53 Setting interface of Auto Trigger
2. Set the delay.
Press Delayto select Auto or Manual mode.
3. Set the number of samples or trigger.
Press Samples/Triggerto set sample count.
4. Set the VMC Out.
Press VMC Outto choose Positive or Negative polarity.