Specifications
All specification applies to 10X probe and All the SDS1000CFL Series Digital
Storage Oscilloscopes.
To verify that the oscilloscope meets specifications, the oscilloscope must first
meet the following conditions:
The oscilloscope must have been operating continuously for thirty minutes
within the specified operating temperature.
You must perform the Do Self Cal operation, accessible through the Utility
menu, if the operating temperature changes by more than 5° C.
The oscilloscope must be within the factory calibration interval
All specifications are guaranteed unless noted “typical.”
Inputs
Input Coupling AC, DC, GND
Input Impedance
1MΩ±2% || 18Pf±3Pf,
50Ω+/-2%(SDS1304CFL/SDS1302CFL,
SDS1204CFL/SDS1202CFL contain this function)
Maximum input
voltage
400V DC+AC PK-PK1MΩ input impedance
X10
),
CAT I
Ch to Ch Isolation
(Both channels in same
V/div setting)
>100:1 at 150MHz (SDS1304CFL,SDS1302CFL)
>100:1 at 100MHz (SDS1204CFL,SDS1202CFL)
>100:1 at 50MHz (SDS1104CFL,SDS1102CFL)
>100:1 at 35MHz (SDS1074CFL,SDS1072CFL)
Probe attenuator 1X,10X
Probe attenuator
Factors Set
1X,5X,10X,50X,100X ,500X,1000X
Vertical System
Vertical Sensitivity 2mV/div -5V/div(1-2-5 order)
Channel voltage offset
range
2mV-100mV: ±800mV
102mV-5V: ±40V