Interval Trigger
The time interval from the rising edge (or falling edge) of input signal passes
through the trigger level to the next rising edge(or falling edge) passes
through the trigger level and the currently set time satisfy the Limit Range (
=,
= , [ --.--], --] [--) selected
In the gure below, the time interval between two continuous rising edge and
the currently set time satisfy the limit range ([ --.--])
Runt Trigger
Runt trigger includes positive runt trigger and negative runt trigger.This mode
is used to trigger pulses that pass through one trigger level but fails to pass
through the other trigger level as shown in the gure below.
High Level
Low Level
Positive runt pulse
Negative runt pulse
In the gure below, the trigger signal are positive runt pulse.